Brand:labtop
Model:iCooler-2009
Name:Low temperature chilled down liquid circulating machine
| Instrument type | Chilled down liquid circulation device |
| Security level (according to DIN12876) | I/NFL |
| Protection Rating protection level | IP20 |
| Temperature range | -20~RT°C |
| DIN12876 temperature stability | 0.5±K |
| Temperature display | LED |
| Temperature control method | Intelligent PID control |
| Temperature transducer | Pt100 |
| External temperature transducer interface | no |
| refrigerant | R404a |
| Cooling Capacity (10 ℃) | 1200W |
| Pump type | Pressure |
| Pump adjustability | no |
| Maximum pump pressure | 1bar |
| Pump maximum flow | 22L/min |
| Pump connection interface | RC1 / 2 |
| Storage tank Volume | 9L |
| Dimensions | 300*500*620mm |
| Temperature display Accuracy | 0.1℃ |
| Light alarm | yes |
| audible alarm | yes |
| Temperature over limit alarm | yes |
| The temperature is not enough alarm | no |
| secondary protection | no |
| Low level alarm | no |
| Calibrating options | yes |
| Instrument fixing method | table |
| Allow continuous operation | 100% |
| Allowable environment temperature | 5..32℃ |
| Allowable Relative Humidity | 80% |
| RS232 interface | option |
| USB interface | option |
| Power | 200. p...230VAC/50HzV |
| Instrument noise | ≤55db |
| Technical parameter standard | DIN 12876 |
Warranty days: 365 days
Non-artificially caused product quality problems, the manufacturer provides free warranty services during the warranty period