| Instrument type | Chilled down liquid circulation device |
| Protection level | I/NFL |
| Protection level | IP20 |
| Temperature range | -20~100℃ |
| Temperature fluctuation | 0.05K |
| Temperature display | TFT |
| Temperature control method | Smart PID |
| Temperature transducer | Pt100 |
| Temperature transducer | without |
| refrigerant | R404a |
| Cooling Capacity 10 ℃ | 1200W |
| Pump type | Pressure/Suction Pumps |
| Pump adjustability | without |
| Maximum pump pressure | 1bar |
| Pump flow | 22L/min |
| Pump connection interface | M16x1 |
| Tank Volume | 9L |
| Dimensions | 300*500*620mm |
| temperature resolution | 0.01℃ |
| Light alarm | Yes. |
| audible alarm | Yes. |
| Temperature over limit alarm | Yes. |
| The temperature is not enough alarm | without |
| secondary protection | without |
| Low level protection | without |
| Calibrating options | Yes. |
| Allow continuous operation | 100% |
| Allowable environment temperature | 5~32℃ |
| Allowable Relative Humidity | 80% |
| RS232 interface | optional |
| USB interface | optional |
| Weight | 35kg |
| Power | 200~230VAC/50Hz |
| Instrument noise | ≤45pcs |
| Technical parameter standard | DIN 12876 |
Warranty days: 365 days
Non-artificially caused product quality problems, the manufacturer provides free warranty services during the warranty period