YN-TQD03 battery separator air permeability Tester is a combination of Gelley method, Shober method, Benteson method three test methods in one machine, suitable for battery separator, film, paper and other air permeability measurement, the instrument includes the following three air permeability measurement methods:
Shober method: Constant differential pressure of 1.00 kPa± 0.01 kPa or 2.50±0.01 kPa, select the appropriate test duration, and determine the air flow through the specimen by test volume.
Geller method: Determination of the time taken to pass through 100 ml of gas at a constant differential pressure of 1.23 kPa.
Benteson method: under a constant differential pressure of 1.47kPa, the air flow through the test surface is recorded after clamping for 5s.
SJ_T 10171-2016 General Test Method for Basic Performance of Alkaline Battery Separators (Air Permeability Part)
GB/T 458-2008 Determination of air permeability of paper and cardboard
GB/T 36363-2018 Polyolefin separator for lithium-ion batteries
| Basic application | Films, flakes | All kinds of plastic films, sheets, separators |
| paper | All kinds of paper, cardboard, composite cardboard | |
| Extend your application | Metal sheets, silicon wafers | Silicon wafers, foils, various metal sheets |
| Corrugated cardboard | Corrugated cardboard, containerboard | |
| textile materials | All kinds of textile materials such as woven fabrics, knitted fabrics, coated fabrics, etc | |
| Nonwovens | All kinds of nonwovens such as diapers, towels, medical masks, etc | |
| Other materials | Solid electrical insulating materials, adhesive tapes, geosynthetics, rubber, etc |
(1) The three test methods of Geerley method, Shober method and Bentzor method are integrated into one machine for users to choose freely to meet the testing needs of different users.
(2) The differential pressure of the instrument is adjustable in the range of 0-3KPa, and the accuracy is up to 0.01kPa, which fully meets the requirements of the test differential pressure of the three methods
(3) The ultra-large test volume of the instrument is 1L, which fully meets the volume requirements of the three methods
(4) The free conversion of the test area of 10.0cm2 and 6.42cm2 is realized by adding a mask
(5) The instrument comes with a temperature and humidity sensor to monitor the temperature and humidity of the Test Chamber in real time.
(6) The instrument has good sealing, and the air leakage does not exceed 1.0ml/h
(7) The instrument comes with a display screen and host computer software, which can be operated offline and connected to the computer for synchronous testing