HUITAI BPHJS-150B programmable high-low temperature Humidity Test Chamber Nominal content volume 150L, Temperature range -40 ℃~ + 150 ℃, Temperature Fluctuation +/- 0.5 ℃, Temperature uniformity ≤ +/- 1.5 ℃, BPHJS-150B programmable high-low temperature Humidity Test Chamber The air duct with built-in baffle fan volute ensures that the air temperature field in the box is uniform and the temperature deviation is small.
Model:
Certificate

HuiTai BPHJS-150B Humidity Test ChamberSPEC

HuiTai BPHJS-150B Humidity Test ChamberDetails

HuiTai BPHJS-150B Humidity Test ChamberPacking list

SKU
NB033546
Studio Volume
150L
Interior size
500*500*600mm
Temperature range
-40~150℃
Temperature Fluctuation
±0.5℃
Temperature Uniformity
≤±1.5℃
Temperature deviation
±1.5℃
Heating rate
1.5℃/min;RT+→200℃≈40min
test method
GB/T 5170.2-2008 temperature test equipment, the calculation method of the average rate of temperature change according to the 8.4.6 of GB/T 5170.2-2008
Humidity Control Range
30~98%R.H
Humidity Deviation
+/- 3% R. H (Humidity Deviation : +/- 5% R. H at ≤ 75% R. H)
Enclosure Material
1.2Mm Advanced Cold Hot Plate Nano Paint
Chamber material
1.0Mm stainless steel plate SUS304 #/brushed plate
Viewing Window
Multilayer Hollow Electric Glass
noise
≤ 72dB (A), one meter away from the front of the box and one meter above the ground for measurement
Refrigeration system
France taikang fully enclosed compressor.
refrigerant
R404A, ozone depletion index is 0
refrigeration method
air-cooled
Display
TH1800
Display
7-Inch display, resolution: 16bit, color touch screen
Display resolution
Temperature 0.1 ℃; Humidity: 0.1% RH; Time 1min
Temperature sensor
Class A Pt100 Armored Platinum Resistor
Memory
SD card
Power supply Voltage
AC380V  50HZ
BPHJS-150B
Humidity Test Chamber
HUITAI BPHJS-150B Programmable Humidity Test Chamber Picture

First, the characteristics of the equipment

1. Design basis and guiding ideology

1.1 This product is an extension of our company's HF series standard products, and similar products have mature applications in electrical and electronics, automobiles, scientific research, military industry, solar energy and other units.

1.2 The technical scheme is prepared according to the technical documents provided by the customer, combined with the installation site conditions and other information.

1.3 Product design and manufacturing to achieve professional technology, stable equipment performance, reliable use, simple operation, convenient maintenance, while considering the economy and implementability of the program.

1.4 Ensure the usability, reliability, professionalism and safety of process equipment;

1.5 Adopt mature and reliable new technologies to reduce energy consumption, save costs and improve quality;

1.6 The selection of accessories adopts the principle of combining domestic (CN) and foreign, domestic (CN) equipment is selected from domestic (CN) familiar manufacturers, and key equipment and electrical components are imported or joint venture products to ensure the professionalism, stability and normal service life of the equipment;

1.7 Electrical control is designed according to the requirements of process and equipment, in line with the relevant national electrical standards, reliable operation, and considering electrical protection and interlocking functions.

2. High performance & low energy consumption

Professional Air Supply Design:

The air duct of the built-in deflector fan volute is used to ensure that the air temperature field in the box is uniform and the temperature deviation is small;

Cooling:

The finned effective evaporator has strong frost resistance, small heat transfer temperature difference and fast cooling rate;

Low power consumption:

Compressor return air cooling circuit and energy regulation circuit, optional "cold end output" type, more effective energy saving.

3. High reliability   

Refrigeration:

3.1.1 The use of French Taikang fully enclosed compressor, complete safety protection functions, high reliability;

3.1.2 The main refrigeration and vacuum accessories are products of international good brands;

3.1.3 Compressor suction pressure adjustment and protection to ensure the safety of the compressor under various Operating Conditions.

Control:

TH1800 "controller, 7 inches" color touch screen, equipped with electrical actuators based on Schneider in France and Fuji in Japan, the system is fully automatic control, and the operation reliability is high.

4. Simple operation

Simple operation: touch screen controller Chinese menu bar display, easy to operate;

Upper computer: optional remote monitoring computer, control system to achieve redundant control and remote monitoring;

The high-precision temperature sensor is adopted, which has high measurement accuracy and small drift.

Real-time display of compressor suction and discharge pressure, easy to access and maintenance;

The internal sound reduction of the unit and the bottom vibration damping design reduce the vibration and noise in the transmission.

The Test Chamber can be designed with test interfaces according to user requirements.

5. Complete functions

Complete test functions: high/low temperature constant temperature test, high-low temperature cycle test.

The host computer remote control, dry air purging (optional) and other functions are complete, which is convenient for users to choose according to the situation.

6. Easy maintenance   

Self-diagnosis: The control system has a self-diagnosis function, when there is overtemperature, compressor overpressure, overload, fan overload and other faults, the controller displays the cause of the fault in Chinese and gives a detailed solution;

Maintenance: The water channel filter is installed in front of the condenser, which is convenient for regular cleaning and prevents the condenser from being dirty and blocked, resulting in a decrease in the cooling rate or even overpressure protection shutdown;

History: Temperature history and alarm records can be viewed on both the touch screen and the optional PC display.

Second, the relevant standards

sequence

project

Contents

1

Satisfy the test method

GB/T 2423.1-2008 Test A: Low temperature test method, IEC68-2-1

GB/T 2423.2-2008 Test B: High temperature test method, IEC68-2-2

GB/T2423.3—2008 Constant damp heat test, IEC68-2-78

GB/T2423.4—2008 alternating damp heat test, IEC68-2-30

GB/T2423.22—2008 test Nb temperature change test

GJB150.3A—2009 High temperature test

GJB150.4A—2009 Low temperature test

GJB150.9A-2009 Damp heat test

GJB573A-1998 301 Temperature and Humidity Test

2

Product manufacturing standards

GB/T10589-2008 Technical conditions for low Temperature Test Chamber

GB/T11158-2008 Technical conditions for high Temperature Test Chamber

GB T5170.1-2008 General provisions for inspection methods of environmental test equipment in electrical and electronic industry

GB/T5170.2 Basic parameter verification method

3. Box structure

The Test Chamber is composed of four parts: insulated box, refrigeration unit, air supply device, control and safety protection system, which adopts an integral design, and can be used after being installed in place and connected to the power supply on site.

1.1 Cabinet

The cabinet adopts an overall upper and lower structure, with the studio in the upper part and the lower part for the unit system.

The side wall of the Incubator equipment from the inside to the outside is: stainless steel inner wall - external stiffener - insulation material layer - decorative shell

1.2 Shell

High-grade hot and cold plate nano baking paint

1.3 Insulation layer

PU (polyurethane) foaming + ultra-fine centrifugal glass wool composite insulation layer with suitable high temperature resistance and thermal insulation effect, with a thickness of 100mm.

1.4 Inner wall

The inner wall of the box is made of 1.0mm SUS304/2B stainless steel plate, and all seams are continuously welded with TIG to ensure that there is no leakage after long-term testing.

The outer wall of the inner box is welded, and the main position is reinforced.

The inner wall bottom plate is designed with drainage holes to discharge the condensate in the box to the outside of the box.

1.5 box bottom

The channel steel at the bottom of the box is welded and formed, and the load-bearing ≥ is 100kg/m².

Cabinet layout

The upper part of the Test Chamber is directly in front of the box gate, and the rear of the box is the air conditioning air supply system.

Air conditioning and ventilation

Side air supply: The air conditioning box is at the back of the studio, and the circulating fan, heater, copper tube aluminum fin evaporator, humidifier, etc. are installed inside.

One set of 0.1HP stainless steel long shaft motor is installed on the top of the box, and the impeller of the stainless steel centrifugal fan is installed on the top of the air conditioning box, and the air is evenly supplied through the air outlet deflector fixed inside the volute to meet the requirements of temperature uniformity.

gate

Single-opening hinged anti-collision door (left hinge, right handle), door opening size 1000W×1000H mm, rotary hinge to ensure that the door opening angle is not less than 120°;

The door hinges are equipped with reinforced hinges to ensure that they can be opened and closed freely.

An anti-frost electric heating belt is installed on the inner edge of the door, and the control system automatically turns on the heating and defrosting according to the temperature of the studio.

Observation window

A set of multi-layer laminated glass observation windows with a viewing area of 300×400 square meters are installed on the door, and the controller automatically turns on the observation window defogging function when the studio temperature is lower than the ambient temperature.

The observation window is composed of pressure-resistant tempered glass + multi-layer hollow electric anti-frost glass.

The electric anti-frost glass is the domestic (CN) advanced technology of conductive coated glass, the observation window has the advantages of high light transmittance, good thermal insulation effect, no resistance wire on the glass surface to affect the observation, when the studio temperature is lower than the ambient temperature, the system automatically opens the observation window defogging function, to ensure that under any Operating Conditions can be clearly observed.

illuminating

One set of 9W lighting fixtures is built into the observation window to ensure that the specimens in the chamber can be clearly observed under any Operating Conditions.

Test holes

1 Φ50mm test hole on the left side of the cabinet, located in the center (customer can specify)

Control cabinets

The control cabinet hanging box is installed at the rear of the box, and the distribution board and exhaust fan are installed inside, and the water and electricity control is isolated in the middle to ensure the safety of the test box.

Fourth, the conditions of installation and use

1. Atmospheric environment

Ambient temperature: 5°C~+35°C, Relative Humidity: ≤85% RH;

Air pressure: 86kPa~106kPa

2. Venue

There is appropriate space for use and maintenance around the equipment, as shown in the figure below:


Detail Figure 1 of Huitai BPHJS-150B Programmable High and Low Temperature Damp Heat Test Chamber

A: Not less than 1000mm

B: Not less than 1000mm

C: Not less than 1300mm

The ground is flat, well ventilated, there is no strong vibration around the equipment, there is no strong electromagnetic field around the equipment, and there are no flammable, explosive, corrosive substances and dust around the equipment.

3. Power supply

AC380V±10%,50Hz±1Hz,;

The user needs to configure the equipment with an independent air or power switch of corresponding capacity in the equipment room of the installation site, and lead the power cord to the air switch of the box, and the lead hole of the power line is at the air switch in the power distribution cabinet of the box.

The grounding resistance is less than 4Ω, and the grounding bolt is behind the right of the box base.

4. Drainage

The drainage hole of the box is behind the left of the box base, and 3 meters of DN10 silicone tube is included.