China

JINGTIAN HBDV-2 Digital Rotational Viscometer with L0 rotor

SE
The HBDV-2 Digital Rotational Viscometer has a reproducibility of ± 0.5% and a measurement accuracy of ± 1%. The measuring range of the HBDV-2 Viscosity meter is 25~ 320000000mPa·s, and it is equipped with seven kinds of rotors.
Model:
Certificate

JingTian HBDV-2+L0 Digital Rotational ViscometerSPEC

JingTian HBDV-2+L0 Digital Rotational ViscometerDetails

JingTian HBDV-2+L0 Digital Rotational ViscometerPacking list

SKU
NB035847
Measuring range
25~320000000mPa·s
Rotor Specifications
1 piece each of rotors L0, R2, R3, R4, R5, R6 and R7
Reproducibility
±0.5%F.S
Measurement Accuracy
±1%F.S
Power
Dedicated power adapter (input 100~240V 50~60Hz)
Rotor speed
0.1~200rpm
Display information
Viscosity, shear rate, shear stress, rotational speed, temperature, percentile torque, rotor number and the maximum Viscosity value that can be measured at the current rotational speed of the selected rotor, etc
Speed control method
CVT
HBDV-2+L0
Digital Rotational Viscometer
JINGTIAN HBDV-2 Digital Rotational Viscometer with L0 rotor Picture

Introduction

The LCD of Shanghai JINGTIAN HBDV-2 digital display rotary viscometer can display viscosity, shear rate, shear stress, rotational speed, temperature, percentile torque, rotor number and the maximum measurable value of the selected rotor at the current rotational speed. Viscosity value etc. It is a digital display rotary viscometer with a measuring range of 320 to 320,000,000 mPa·s.

JINGTIAN HBDV-2+L0 Digital Display Rotary Viscometer Details Figure 1

Introduction

Accurately measure ultra-low viscosity samples

Sample volume only needs 25~30mL

It can measure lower viscosity and be used with different types of viscometers

Details of JINGTIAN HBDV-2+L0 Digital Display Rotary Viscometer 2

FAQ
QColloidal, viscosity range 10,000 or so, this one is OK?
A What kind of glue? Glues are all non-newtonian fluids, which are related to Measurement conditions. You need to ask a little more carefully.
QThe same rotor test liquid viscosity, but set different Rotation speed measurement of the same liquid, the measured viscosity is different What is the reason?
A This is the reason for non-Newtonian liquids. Rotation speed, stator and rotor, temperature, and some special samples also need to determine the measurement time. This requires users to refer to industry measurement standards
QDo you have the Viscometer involved in ISO12634?
A The HBDV-2 + L0 Viscometer will do
QIs there any application difference between RVDV and HBDV? It is used on colloids.
A RVDV can measure sealant, silicone, nail-free glue, etc., HBDV can measure silicone rubber, Glass glue and other more viscous
QWhich Viscometer displays shear rate and shear force?
A The DV-2 series has shear rate and shear force display. The DV programmable download software can program the shear rate change of the Viscometer in advance, and then collect and draw data through the data collection software
QI haven't done any testing before, and I don't know how to do it. How to choose a Viscometer?
A It is easy to make the results of the viscosity test. After the measurement, he said that it is different from others, so it is wrangling. If the next home has viscosity requirements for his products, ask the next home for detection methods. If the purchase inspection, ask the previous home for Measurement conditions and methods. There are many viscosity measurement methods, and the measurement results cannot be converted.