| test space | ≥ 0.5 cubic meters with Glass observation door |
| combustion gas | Butane gas Butane |
| Gas Connectivity | Quick Connector with self-locking function |
| flame height | 12mm +/- 1mm adjustable |
| Thermocouple | Imported from the United States OMEGA, diameter 0.5mm (optional) |
| Flame temperature requirements | The time from 100 ℃ to 700 ℃ is within 23.5 +/- 1 S |
| Timer | Adjustable from 0 to 99 minutes and 99 seconds, combustion time can be set, and combustion duration can be recorded |
| Control mode | PLC + touch screen |
| Export/Print Function | USB data export function (optional: after the test is completed, the results can be printed out through a thermal printer) |
| System protection | When the touch screen is in normal use, once a function is activated, the Miscellaneous function will be disabled to prevent mutual interference |
| insulation grade | Class F |
| Protection class | IP20 (suitable for indoor use) |
| Fan noise | 65 decibels |
| Chill down, forced air cooling | Side wind, top wind |
| Operating Mode | Can work continuously |
| protection function | Gas alarm, beep prompt, overcurrent protection, etc |
| Applicable environment temperature | -10℃~+50℃ |
| Mobile mode | Adopt load-bearing universal casters, 2 at the front and rear |
| Applicable Relative Humidity | ≤95% |
| altitude | ≤ 2500 meters |
| Power | AC 220V/50H |
Warranty days: 365 days
1. For product quality problems caused by non-human products of instruments and equipment, the manufacturer will provide free warranty services during the warranty period;
2. There is no warranty period for fragile and consumable parts, and if you find product quality problems after receiving the goods, please apply for return and exchange within 7 days, and you will not wait for the overdue ones;