| Interior size | Inner box size 5 +/- 0.02 m ³; width 1600mm depth 1600mm height 2000mm |
| Temperature & Moisture | Temperature range: 20 ℃~ 65 ℃; humidity range: 30%~ 75% RH (at + 20~ + 60 ℃) |
| Instrument resolution | Temperature: 0.1 ℃; Humidity: 0.1% RH; |
| Temperature Fluctuation | ±0.2℃; |
| Temperature Uniformity | ≤1.5℃ |
| Humidity stability | ≤±1.5%RH |
| air displacement rate | 1~ 3 times/hour (adjustable) |
| Flow rate Accuracy | Digital Flow rate meter, Accuracy class 0.5; 4-40m ³/L |
| Air tightness | When the pressure is over 1000pa, the leakage is less than 10- 3 × 1m 3/min. |
| Stabilization time | Temperature Stabilization time is less than 40min, Humidity Stabilization time is less than 60min, and the equipment can run continuously for not less than 30 days; |
| background indicator | TVOC≤0.05 mg/mm³ |
| noise | ≤ 60 decibels (A sound level), there will be no resonance sound between the box door and the equipment door |