- SKU
- NB039408
- mV range
- -1999mV ~ 1999mV
- pH Measurement range
- (-2.00 ~ 19.99)pH
- TDS range
- 0~100g/L
- Measurement parameters
- pH, mV, Conductivity Detector, Resistivity, Salinity, TDS
- PH resolution
- 0.1/0.01pH
- pH measurement Accuracy
- Electric meter: ± 0.01pH, matching: ± 0.02pH
- PH input current
- ≤2×10(-12) A
- PH input impedance
- ≥ 1 × 10 (12) Oh
- PH stability
- ±0.01 pH/3h
- Temperature compensation range
- (0~100) ℃ (automatic/manual)
- MV resolution
- 1mV
- mVAccuracy
- ±0.1%FS
- Conductivity Detector Measurement range
- (0.00~19.99)μS/cm、(20.0~199.9)μS/cm、(200~1999)μS/cm、 (2.00~19.99)mS/cm、(20.0~199.9)mS/cm
- Salinity range
- 0~100ppt
- Resistivity Measurement range
- 0 ~ 100MZ · cm
- Conductivity Detector resolution
- 0.01/0.1/1μS/cm 0.01/0.1mS/cm
- Conductivity Detector measurement Accuracy
- Electric meter: ± 1.0% FS, matching: ± 1.5% FS
- Temperature compensation range
- (0~100) ℃ (automatic/manual)
- Electrode constant
- 0.1 / 1 / 10 cm-1
- reference temperature
- reference temperature
- data storage
- 150 groups
- Save content
- Numbers, measurements, Measurement Units and temperature values
- Power
- 2 AA batteries (1.5V × 2)
- Dimensions
- 165*90*32mm
- Weight
- 310g
- Environment temperature
- 5~ 35 ℃ (0.01 grade)
- Environment humidity
- ≤85%
- Protection class
- IP57
- Mass and safety certification
- ISO9001:2000,CE和CMC
P613
P6 series acidity meter/conductivity meter
Acidity + Conductivity Detector two-in-one, automatic identification of 13 kinds of pH standard buffer solution, high purity water and ammonia pure water 2 kinds of pH measurement mode, auto calibration, automatic temperature compensation.
Features
1. Large-screen LCD display with controllable white backlight.
2. Humanized design: bracket design, convenient for customers to use. Anti-skid glove design, convenient for customers to operate.
3. Dual-parameter instrument, equipped with pH electrode and conductivity electrode, the instrument automatically recognizes the electrode.
4. Built-in microprocessor chip, with intelligent functions such as automatic calibration, automatic temperature compensation, data storage, function setting, self-diagnosis information, automatic shutdown and low Voltage display.
5. Automatically recognize 13 pH standard buffer solutions, there are three series of standard buffer solutions to choose from: European and American series, NIST series and Chinese series.
6. The pH measurement mode of high-purity water and the pH measurement mode of pure water with ammonia can be set.
7. Automatically identify 8 kinds of conductivity standard solutions, and there are two series of standard solutions to choose from: European and American series and Chinese series.
8. Using excellent conductivity measurement technology, using a conductivity electrode with a constant K=1 requires only one point of calibration to meet the measurement accuracy requirements of 0.5μS/cm ~ 200 mS/cm.
9. Switchable conductivity, TDS, salinity and resistivity. TDS and salinity are calculated using polynomials to ensure the conversion accuracy of the full range.
10. Conductivity measurement mode has high-purity water non-linear temperature compensation function.
11. The instrument complies with IP57 waterproof level, equipped with calibration solution and suitcase.
Packing list:complete-built unit X1, manual X1, certificate of conformity X1, warranty card X1
[Note]Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.