Yoke DDS-307A desktop conductivity meter 1.0 grade

SE
Instrument level 1.0, automatic/manual temperature compensation, dual-row digital, liquid crystal display.
Model:
Certificate

Yoke DDS-307A Desktop conductivity meterSPEC

Yoke DDS-307A Desktop conductivity meterDetails

Yoke DDS-307A Desktop conductivity meterPacking list

SKU
NB039411
Instrument level
Level 1.0
Measuring range
( 0 ~ 1,999 ) μS ; ( 2.0 ~ 19.99 ) μS ; ( 20.0 ~ 199.9 ) μS ; ( 200 ~ 1999 ) μS ; ( 2.0 ~ 19.99 ) mS ; ( 20.0 ~ 199.9 ) mS
Measuring range
-5~105℃
Electronic unit intrinsic error
Electronic unit reference error: ± 1.0% FS; temperature coefficient error of indication: ± 1.5%/℃
temperature compensation coefficient
2.0(%)
Temperature compensation range
0~100℃
Power
AC(220±22)V;(50±1)Hz
DDS-307A
Desktop conductivity meter
Instrument level 1.0, automatic/manual temperature compensation, dual-row digital, liquid crystal display.
Yoke DDS-307A desktop conductivity meter 1.0 grade Picture

Features

1. Large screen, blue backlight, double rows of numbers, liquid crystal display.

2. With manual and automatic temperature compensation function.

3. With 0~10mV DC output signal.

4. Equipped with DJS-1C conductivity electrode (platinum black) T-818-B-6

5. Type temperature sensor.

FAQ
QHow is the Conductivity Meter calibrated with standard liquid?
A The electrode is immersed in the standard solution and the temperature of the standard solution is controlled at 25 +/- 0.1 ° C. Standard solution selection: electrode coefficient 0.1 can be used for standard solution of 146.6 μs/cm, electrode coefficient 1.0 can be used for standard solution of 1408 μs/cm, and electrode coefficient 10 can be used for standard solution of 12.85ms/cm or 111.3ms/cm. The current conductivity measured standard solution result K is known, the standard solution conductivity is 1408μS/cm, and the re-calibrated electrode Changshu J is unknown. J known/K known = J unknown/1408