China

Yoke YG-01 button desktop dissolved oxygen analyzer, fluorescence method

SE
Fluorescence dissolved oxygen analyzer, range 0-20.00ppm, concentration 0-200%, with fluorescence dissolved Oxygen electrode, electrode without film, no electrolysis solution, no polarization.
Model:
Certificate

Yoke YG-01 Button type desktop Dissolved Oxygen Meter (fluorescence method)SPEC

Yoke YG-01 Button type desktop Dissolved Oxygen Meter (fluorescence method)Details

Yoke YG-01 Button type desktop Dissolved Oxygen Meter (fluorescence method)Packing list

SKU
NB039444
Measurement parameters
Concentration, saturation, temperature
Display
6.5 inch LED screen
DO range
0.00~20.00mg/L(ppm)
DO resolution
0.01/0.1
DO measurement Accuracy
Electric meter: ± 0.10mg/L; matching: ± 0.40mg/L
DO measurement Accuracy
±1%
DO% Measurement range
0.0~200%
DO% resolution
0.1%
DOAccuracy
±2.0%FS
Temperature Measurement range
(0~100)℃
temperature resolution
0.1℃
temperature relative Accuracy
±0.1℃
Temperature compensation range
(0~100) ℃ (automatic/manual)
Salinity compensation range
0~45ppt (manual)
Air pressure compensation range
0~200kPa (automatic)
Barometric resolution
0.1
Air pressure Accuracy
± 2%
Stability
(±0.07mg/L)/1h
Response time
≤10s
data storage
2000 groups, external USB for Mass storage
Communication interface
USB, wireless Bluetooth (optional)
Environment temperature
5~40℃
Power
12V1A (110-220, 50Hz-60Hz adapter)
Dimensions
160*190*70mm
Weight
880g
Working Relative Humidity
5~85%
YG-01
Button type desktop Dissolved Oxygen Meter (fluorescence method)
Fluorescence dissolved oxygen analyzer, range 0-20.00ppm, concentration 0-200%, with fluorescence dissolved Oxygen electrode, electrode without film, no electrolysis solution, no polarization.
Yoke YG-01 button desktop dissolved oxygen analyzer, fluorescence method Picture

Features

1. It adopts 6.5-inch LED screen with complete and intuitive parameter display and user-friendly interface design, which is convenient for users to operate and use.

2. Omron long-life light-touch button is adopted, which is convenient and durable, and can reach more than 100,000 times.

3. Meet the GLP requirements of international standards, with intelligent functions such as automatic calibration, automatic temperature compensation, data storage, USB output, clock display, wireless printing, function setting and self-diagnosis information.

4. The built-in large-capacity storage can store 2000 sets of measurement data, the data can be stored and transferred through the U disk, and the stored data can be opened by the general office software EXELL, which is very convenient for customers to edit.

5. The dissolved oxygen measurement mode has the functions of automatic temperature compensation, automatic salinity compensation and automatic air pressure compensation.

6. The instrument can be equipped with a Bluetooth module, which supports wireless Bluetooth printing, which is convenient for users to operate.

7. Equipped with a new type of dissolved oxygen electrode with a temperature sensor, the dissolved oxygen measurement mode has the functions of automatic temperature compensation, manual salinity compensation, and manual air pressure compensation, which is more convenient to use and more accurate to measure.

8. Fluorescent dissolved oxygen sensor, the electrode has no membrane, no electrolyte, no polarization. The electrode does not need to consume oxygen, is not affected by the flow rate, is free from maintenance, has a long service life, and is easy to replace the fluorescent cap.

9. Comply with IP54 dust-proof and splash-proof design.

FAQ
QThis execution standard
A JJG 291-2018
QIt is written above that it is not affected by Flow velocity. Do you not need Stirrer?
A Fluorescence does not consume oxygen and does not require stirring
QWill bubbles in the sensitive film affect the test results when the dissolved oxygen meter is tested?
A Yes, the surface of the sensitive film can not accumulate bubbles, otherwise it will affect the Test accuracy.
QDoes Sensors need to be cleaned every day?
A No, wash it every 30 days.
QHow long does it take to replace the fluorescent cap?
A It is recommended to replace it once a year.
Knowledge
Standard