China

LEICI PHS-3E precision pH meter 0.01 level pH value mV (ORP) temperature value

SE
0.01 level; desktop structure; pH value, mV (ORP), temperature value; automatic temperature compensation; two-point calibrating; automatic identification of standard buffers.
$459.00/set
PromotionLoading...
Coupon
Model:

LeiCi PHS-3E pH meterSPEC

LeiCi PHS-3E pH meterDetails

LeiCi PHS-3E pH meterPacking list

SKU
NB004024
Instrument level
Level 0.01
Measurement range mV
-1999~1999mV
mV resolution
1mV
mV electronic unit error of indication
±0.1%FS
PH range
(-2.00~18.00)pH
pH resolution
0.01pH
An indication of pH electronic unit error
±0.01pH
Temperature range
(-5.0~110.0)℃
Temperature Resolution
0.1℃
Temperature electronic unit error of indication
±0.2 ℃
Power Supply
Power adapter (input: AC100~240V, output: DC9V)
Dimensions
242*195*68mm
Weight
0.9kg
PHS-3E
pH meter
0.01 level; desktop structure; pH value, mV (ORP), temperature value; automatic temperature compensation; two-point calibrating; automatic identification of standard buffers.
LEICI PHS-3E precision pH meter 0.01 level pH value mV (ORP) temperature value Picture

Introduction

The PHS-3E precision PH Meter is a desktop digital display PH meter, both models have a measurement resolution of 0.01pH and a mV value resolution of 1mV.The PHS-3E precision PH Meter is automatically temperature compensated and equipped with E-201F composite electrode and respectivelyE-301F composite electrode.The PHS-3E precision PH Meter measures temperature values. This precision PH Meter is a PH meter used in the laboratory, which is widely used in chemical industry, food, scientific research institutions, colleges and universities and other fields.

Features and advantages

Features

● High-definition LCD display, button operation

● Support functions such as automatic shutdown, power-off protection and factory reset

● Support IP54 protection level

● Triple composite pH electrode, electrode holder, dust cover and calibration buffer powder are standard

 

Intelligent detection and automatic identification

● Intelligent endpoint discrimination, support balance measurement mode and continuous measurement mode

● Support automatic/manual temperature compensation

● Support 1-3 points of pH electrode calibration

● Automatically identify three pH standard buffer solutions: GB 4.00pH, 6.86pH and 9.18pH, and support custom pH buffer solutions

 

Data management

● Support data storage (50 sets), viewing, and deletion


Installation method:

1. Unplug the electrode base bracket, unscrew the nut at the lower end of the electrode bracket, and willThe PHS-3C/3E precision pH meter electrode holder is installed to the end and the nut is tightened.

2. Install the power supply and plug it into the socket.

3、After the PHS-3E precision PH Meter is connected to the power supply, you can see that 0mV is displayed, and generally this value does not exceed 0+0.2mV.

4. Install the electrode on the electrode holder, unplug the short-circuit plug, and connect the PH electrode.

Troubleshooting and solutions

1. When measuring or calibrating, if the data is not accurate, you can unplug the electrode and plug it in with a short circuit. If the number is 0±0.2mV, it meansThe PHS-3C/3E precision PH Meter is normal.

2. If the calibration data is not accurate, you can press and hold the "confirm" button on the measurement status for more than 3s.PHS-3C/3E precision PH Meter displays "SYS rSt", release the "Confirm" button, the PH meter flashes, and then press the "Confirm" button.The PHS-3C/3E precision PH Meter calibrates the slope of the reset electrode.

Precautions for use

1、The PHS-3E precision PH Meter is a precision instrument with a high input impedance. When in use, keep the input port dry and clean. When the PH meter is not in use, insert the Q9 short-circuit plug into the socket to prevent moisture and dust.

2、The PHS-3E precision pH meter and its power supply should be well grounded.

3. When measuring, the electrode lead-in line should be kept in a stationary state.

4. When calibrating, the wrong solution should not be matched, otherwise it will be causedThe PHS-3C/3E precision PH Meter is incorrect.

FAQ
QWhat kind of electrode is PHS-3E?
A E-301-F
QWhat's the difference between these 3? I think the parameters are the same.
A 3C: range -2-18ph, manual temperature compensation, 3-point calibration, 50 sets of results are stored. 3E: range -2-18ph, manual + automatic temperature compensation, 3-point calibration, 50 sets of results are stored. 3F: range -2-20PH, manual + automatic temperature compensation, 5-point calibration, 500 sets of results are stored, there is RS-232 interface to connect to PC, bilingual switching in Chinese and English
QCan this pH meter measure electrode potential?
A Hello, if this PHS-3E wants to test the electrode potential of the electrode, it needs to be equipped with an ion selective electrode.
QWhat are the three interfaces on the left behind the PHS-3E acidity meter?
A Hello, there are six interfaces behind the acidity meter, from left to right, namely measurement electrode socket, reference electrode interface, temperature electrode socket, fuse, power switch, power socket.
QThe pH meter is used for the first time, and the test data is not accurate. What is the reason?
A Hello, the PHS-3E pH meter needs to be soaked in 3mol/h potassium chloride solution for 24h after the first use or reuse after a long time.
QHello, what is the difference between the E-201F/E-301F composite pH electrode of this pH Meter?
A Hello, the 201 electrode pH Meter can't test the temperature value; PHS-3E supports automatic temperature compensation, which can measure pH value, mV (ORP) and temperature value. PHS-3C: Only pH and mV can be measured.
Knowledge
Standard