China

TONGAO TR 6900Z Desktop heavy metals STIP-scan

SE
Detection of heavy metals content, including total chromium, hexavalent chromium, copper, zinc, nickel, iron and other parameters, built-in thermal printer, using digestion Colorimetric integrated tube Colorimetric.
$3149.00/set
PromotionLoading...
Coupon
Model:
Certificate

TongAo TR 6900Z Desktop STIP-scanSPEC

TongAo TR 6900Z Desktop STIP-scanDetails

TongAo TR 6900Z Desktop STIP-scanPacking list

SKU
NB040457
measurement error
≤± 5%
Repeatability
≤±5%
Optical inspection Stability
≤ +/- 0.001A/20min (100,000 hours life)
Data storage
30,000 records (including item, date, time, sample ID, etc.)
Linear dispersion quantity
96 Linear dispersions (expandable)
printer
Built-in thermal printer
Display
5-Inch high-definition color LCD
batch Volume
16/25 water samples/batch
Colorimetric method
Digestion Colorimetric Tube
Sample compatibility
16 mm (round), 25mm (round)
Communication interface
USB cable, LAN, etc
exception notification
Blank comparison prompt, overrange prompt, etc
Operating interface
Membrane push-button operation, Chinese/English (optional)
Power Supply
AC(220V±10%),50Hz
Power
5W
Protection level
IP30
Relative Humidity
≤ 85% non-condensing
Ambient Temperature
5~40℃
Host size
348*250*145mm
Instrument Weight
Main unit less than 2.6kg
Total chromium detection method
GB7466-87
Measurement range of total chromium
0~5mg/L
Total chromium detection lower limit
0.01mg/L
Total chromium digestion environment
125℃,30min
Hexavalent chromium detection method
GB7467-87
Measurement range of hexavalent chromium
0~5mg/L
Lower limit of hexavalent chromium detection
0.01mg/L
Hexavalent chromium digestion environment
No need to digest
Copper detection method
Diacetaldehyde oxalyl dihydrazone
Measurement range
0~50mg/L
Lower limit of copper detection
0.01mg/L
Copper digestion environment
No need to digest
Nickel detection method
GB11910-89
Measurement range
0~4mg/L
Lower limit of nickel detection
0.01mg/L
Nickel digestion environment
No need to digest
Zinc detection method
Xylenol Orange Spectrophotometry
Measurement range
0~15mg/L
Zinc detection lower limit
0.01 mg/L
Zinc digestion environment
无需消解
Iron detection method
HJ/T345-2007
Measurement range
0~50 mg/L
Lower limit of iron detection
0.01mg/L
Iron digestion environment
No need to digest
TR 6900Z
Desktop STIP-scan
Detection of heavy metals content, including total chromium, hexavalent chromium, copper, zinc, nickel, iron and other parameters, built-in thermal printer, using digestion Colorimetric integrated tube Colorimetric.
TONGAO TR 6900Z Desktop heavy metals STIP-scan Picture

Introduction

The all-in-one heavy metal Detector can quickly measure total chromium, hexavalent chromium, copper, nickel, zinc, iron, etc. in water. It integrates mature technologies in the field of water quality analysis, making the entire analysis system more complete, and the operation is simpler and faster. More stable and reliable. Widely used in emergency monitoring, industrial wastewater, domestic (CN) sewage, sewage treatment, pharmaceutical and chemical industry, food, beverage, textile printing and dyeing, electroplating, papermaking, petroleum metallurgy, river management, scientific research units, colleges and universities and other fields.

operating principle

The integrated digestion colorimetric tube and the multi-channel water quality rapid analyzer can be used seamlessly, and are compatible with φ25mm colorimetric tubes.Tongao Technology TR 6900Z Desktop Water Quality Detector Details Figure 1

● The detection of each parameter conforms to national standards, industry standards or international standards.

●  Built-in two COD detection methods

Method 1: HJ/T399-2007 (national standard)

Method 2: Import detection method

●  The standard detection parameters are as follows:

- COD (premade reagents)

- COD (powder pack reagent or custom reagent)

- Ammonia nitrogen (NH3-N)

- Total phosphorus (TP)   

- Total Nitrogen ( TN )

- heavy metal

●  Strong scalability in the later stage, including but not limited to the following parameters:

- Phosphate - Manganese method COD 

- Nitrite - Nitrate Nitrogen

- Free/Total Chlorine - Fluoride

 …

Principle

1. Digestion colorimetric integrated tube

The digestion and colorimetric integrated tube is used for operation, and the digestion of the sample and the reading of the result share a reaction tube, and there is no need to transfer the liquid to the cuvette for detection.

2. Display the result directly

A cold light source with a specific wavelength is used to measure the absorbance of the sample, and the instrument system converts the detection result directly into the sample concentration value through a specific algorithm. 

3. Measure multiple times and remove outliers

Measure the water sample 10 times or more, and take the average value after removing outliers, so as to reduce the influence of dirt, scratches, fingerprints or defective glass tubes on the results.

4. Detection of abnormal prompts

Prompts for abnormal detection, such as prompts for blank comparison, prompts for results exceeding the upper limit of the range, etc.


FAQ
QIs the price listed on the website the final cost for overseas purchases?
ANo, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
QWhat is the specific process for purchasing the product?
AFirst, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
QWhat should I do if the instrument I purchase needs repair?
AAfter-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.
Standard