Brand:Moderner
Model:XTH-080C
Name:High-low temperature alternating (heat and humidity) Test Chamber
| Power | AC220V 50Hz |
| Maximum current | 25A |
| Maximum Power | 16.2KW |
| Temperature control method | Equilibrating Conditioning (BTC) |
| Temperature range | -70~150℃ |
| Humidity control range | 20~98%R.H |
| Temperature Fluctuation | ≤±0.5℃ |
| Temperature Uniformity | ≤2℃ |
| Temperature deviation | ≤±2℃/+2,-3%RH |
| Heating rate | -70~+100℃≤60min;-70~+150℃≤75min |
| Cooling rate | ﹢20~-70℃≤80mins |
| Enclosure Material | Cold rolled steel plate |
| Chamber material | SUS304 2B stainless steel plate |
| Insulation material | Rigid Polyurethane Foam (+ Ultrafine Glass Cotton) |
| refrigeration method | Single compression refrigeration (air-cooled) |
| Refrigeration system | Imported fully enclosed compressor |
| Compressor Capacity | 6+6HP |
| Condenser | Heat sink type chilling down (double as dehumidifier) |
| Water supply method | Electromagnetic pump |
| Water tank Capacity | 20L |
| Heating system | Nickel chromium wire heating |
| blower | centrifugal fan |
| Temperature transducer | High-precision platinum resistor (PT100) |
| Controller | UMC1200 programmable touch LCD Controller with remote monitoring and debugging software package |
| Interior Volume | 800L |
| Interior size | 1000*1000*800mm |
| Dimensions | 1460*2180*1240mm |
| Weight | 715kg |