As the air source of the gas chromatograph, the cold-dry hydrogen generator can replace the high-pressure cylinder in the laboratory, making the analysis work more convenient and safer. Equipped with a cold-drying system, the generated gas is condensed and adsorbed, eliminating the need to frequently replace the color-changing silica gel.

⪠7-inch color display screen, online monitoring of cold-drying temperature, output pressure, and output flow, clear cold-drying status, electrolysis status, water level status, and online leakage detection at a glance
⪠The output pressure and the pressure inside the tank adopt precision pressure sensors and are digitally displayed on the screen. There is no need for traditional pointer pressure gauges. It has a normal pressure starting device, which extends the life of the compressor.
⪠With automatic water replenishment function, it controls the pump for liquid replenishment without manual replenishment.
⪠The output gas temperature is low, better than -70℃ dew point, and the output gas has low moisture content
⪠It has the memory function of the current startup time and accumulated working time, which facilitates humanized management of the laboratory.
⪠Use low-temperature cold drying and pressure swing adsorption water removal technology to avoid frequent replacement of discolored silica gel and reduce the workload of laboratory personnel
⪠The purification tube is made of high-strength PC material and is molded in one injection molding. It is resistant to reagents, high temperatures and impact.
⪠Easy to operate, just turn on the power switch to produce gas. The output pressure is stable and can be used by multiple chromatographs
⪠Add potassium hydroxide solution for the first time, then add distilled water for continuous use
⪠The instrument is equipped with a hydrogen leakage protection function. When hydrogen leaks, the instrument automatically cuts off the power supply and alarms.
⪠The instrument has water protection function to ensure that the gas chromatograph is not damaged
⪠Equipped with a non-return check valve to avoid liquid return.
⪠Replace traditional high-pressure hydrogen cylinders and instrument the laboratory
⪠Integrated fully automatic operation of cold drying, water replenishment, electrolysis and oxygen exhaust, truly achieving 24-hour unattended operation