Back

Apkjgroup 3AP-CJ-100C Thermal Shock Test Chamber Specifications

Search
Aipei 3AP-CJ-100C air-cooled Thermal Shock Test Chamber -65 ℃

Brand:Apkjgroup

Model:3AP-CJ-100C

Name:Thermal Shock Test Chamber

DetailsContact
Specifications
Temperature range -65 ℃~ + 150 ℃ (air-cooled)
Temperature Fluctuation ±0.5℃
Temperature deviation ±2.0℃
Weight Approx. 750kg
Inner Volume 100L
Liner size 500*500*400mm
Dimensions 1500*2000*1650mm
Low Temperature Impact Range -10 ℃~ -65 ℃ (optional)
High Temperature Impact Range 60 ℃~ 150 ℃ (arbitrarily adjustable)
Test environmental conditions Environmental temperature around the machine is maintained between + 25~ + 30 ℃, Relative Humidity: ≤ 85%; air pressure: 86kPa~ 106kPa normal atmospheric pressure.
High Temperature Impact Range RT~+165℃
Heating rate +50℃→+165℃ ≤30min
Low Temperature Impact Range RT~-75℃
Cooling rate +20℃ → -75℃≤≈60min
Test method Using heat storage, cold storage structure, strong cooling hot air circuit switching mode into the test area, Impact high temperature zone or low temperature zone temperature into the test area Impact, complete the hot and cold temperature Impact test.
temperature recovery time ≤ 5min; the conversion temperature only needs ≤ 10s (throttle opening time: within 3 seconds)
Impact exposure time ≥30min
Enclosure Material The outer material can be selected from SUS304 good stainless steel or fine powder paint treatment
Liner material Stainless steel plate (SUS #304)
Box insulation material Rigid polyurethane foam
Air conditioning channel Fan, heating device, evaporator, damper, temperature sensor
Specimen shelf 1 set, a total of two layers (with four hooks) bearing 25KG (please indicate when placing an order for larger bearing)
Lead Hole (Optional) 1, located on the left side of the box, with a round hole with a size of 0019 50mm
Test chamber door Single open, left hinge, right handle
Circulating Fan High temperature/ambient temperature exposure: 3φ, 400W, centrifugal; low temperature exposure: 3φ, 400W, centrifugal
Control Panel Controller display, fault indicator, timer, USB or R232 interface, over-temperature protection setter, main power switch
Mechanical room Refrigeration unit, connected drainage device, exhaust fan
Distribution control cabinet Total power leakage circuit breaker, Controller, distribution board, heat dissipation fan, high and low temperature Circulating Fan motor
Heating system Nickel chromium alloy heating wire heating device
Control mode of heating No touchpoint equal period pulse width modulation, SSR (solid state relay)
Power cord holes and drainage holes Located on the back of the box
Refrigeration system France Taikang or Japan Hitachi fully enclosed low noise rotary compressor
Evaporator finned tube heat exchanger
Condenser shell and tube heat exchanger
Throttle device Thermal expansion valve, capillary tubing
Evaporative condenser Stainless steel brazed plate heat exchanger
Control mode of refrigerator Control system PLC (programmable logic Controller) automatically adjusts the operating conditions of the refrigerator according to the test conditions, compressor return air chill down circuit, energy regulation circuit, condensing pressure control valve
refrigerant R404a/R23
Controller 7 inch large touch AP-950 programmable temperature Controller full touch screen Controller
monitor LCD liquid crystal display
operation mode procedural approach
setting method All Chinese LCD display (Chinese and English can be switched) LCD touch panel (resistant to 100.000 line drawing test)
program Capacity Up to 120 groups, each program is divided into two or three zones, two zones have 2 segments, and three zones have 3 segments. Each command can be up to 999 times; the program segment time is set to 0~999Hour59Min
Display resolution Temperature: 0.1 ℃; Time: 0.1 min
electrical input Thermocouple
storage device USB and R232 communication interface
Control mode Anti-integral saturation PID, fuzzy algorithm; BTC equilibrating temperature control mode
Curve recording Real-time display of graphics during program execution Linear dispersion
subsidiary function It has power-off program memory, automatically starts and continues to execute program functions after power-back, with date, time adjustment, appointment start, shutdown function and screen lock (LOCK) function.
Environmental conditions Temperature: 5 ℃~ 35 ℃; Relative Humidity: ≤ 85%; air pressure: 86kPa~ 106kPa normal atmospheric pressure.
air source pressure (5~7)kg/cm2
Air source connecting pipe diameter Φ8mm
Flow rate of air source 0.1m3/min
Maximum Power Approx. 10.0KW [Voltage range: no more than +/- 10%]
Storage environment When the equipment is not working, the temperature of the environment should be kept within 0 ℃~ + 45 ℃; when the Environmental temperature is lower than 0 ℃, the water remaining in the equipment should be discharged clean to avoid the water in the pipeline freezing and damaging the pipeline.
Maximum current 15A
Packing list
machine X1, manual X1, certificate X1, warranty card X1
After-sales instructions

Warranty days: 365 days

[Note] For after-sales services outside the Chinese mainland, if the instrument requires repair, the shipping and possible associated taxes shall be borne by the sender. Generation address of the manual: http://en.nbchao.com/p/42369/Specification.html , which is only used by users who purchase commodities from the NBCHAO.