China

HUITAI GDT-150BH High and low temperature alternating climate chamber with Humidity control 150L/-40~ 150 ℃

SE
Temperature range -40~ 150 ℃, Humidity range 20% R.H~ 98% R.H, Studio volume 150L, Voltage 220V/25A, Power 4.5Kw, Inner chamber dimensions 500 * 500 * 600mm.
Model:
Certificate

HuiTai GDT-150BH High and low temperature alternating climate chamberSPEC

HuiTai GDT-150BH High and low temperature alternating climate chamberDetails

HuiTai GDT-150BH High and low temperature alternating climate chamberPacking list

SKU
NB044240
Temperature range
-40~150℃
Temperature Fluctuation
≤±0.5℃
Temperature deviation
≤±2℃
Temperature Uniformity
≤2℃
Temperature Resolution
0.1℃
Heating rate
-40 ℃ → 100 ℃ 3~ 5 ℃/min (mechanical refrigeration, under standard load)
Cooling rate
+ 25 ℃ → -40 ℃ 1~ 1.5 ℃/min (mechanical refrigeration, under standard load)
Humidity range
20% R. H~ 98% R. H (see Temperature Humidity Controllable Chart)
Humidity Deviation
≤±3%R.H
Humidity stability
±1.5%R.H
Humidity resolution
0.1%R.H
Standard load Capacity
1Kg aluminum sheet, 200W calorific value
Interior Volume
150L
Voltage
220V/25A
Power
4.5Kw
Interior size
500*500*600mm
Dimensions
700*1400*1700m
GDT-150BH
High and low temperature alternating climate chamber
Temperature range -40~ 150 ℃, Humidity range 20% R.H~ 98% R.H, Studio volume 150L, Voltage 220V/25A, Power 4.5Kw, Inner chamber dimensions 500 * 500 * 600mm.
HUITAI GDT-150BH High and low temperature alternating climate chamber with Humidity control 150L/-40~ 150 ℃ Picture

Instrument interface

GDT series configuration

temperature range

A:-20~150

B:-40~150℃

C:-60~150℃

Humidity range20%~98% Relative Humidity (model with H)
Display instrument7-inch TFT color LCD display (can be connected to RMCS control software)
Operation modeFixed value mode, program mode (preset 100 groups, 100 steps, 999 cycles)
control methodBTC balance temperature control method + DCC (intelligent cooling capacity control) + DEC (intelligent electrical control) (temperature test equipment)
BTHC balance temperature and humidity control method + DCC (intelligent cooling capacity control) + DEC (intelligent electrical control) ( Temperature and humidity testing equipment)
Curve recording functionIt has battery-protected RAM that can save the device’s setting values, sampling values ​​and sampling time; the
maximum recording time is 350 days (when the sampling cycle is 1/min)
Software usage environmentThe upper computer operating software is compatible with XP, Win7, Win8, and Win10 operating systems (provided by the user)
Communication functionChoose one of RS-485 interface MODBUS RTU communication protocol and Ethernet interface TCP/IP communication protocol; support secondary development and
provide host computer operation software, RS-485 interface single device link, Ethernet interface can realize remote communication of multiple devices Online.
Way of workingA/B: Mechanical single-stage compression refrigeration system
C: Two-stage cascade compressor refrigeration method
way of observationHeated viewing window with LED internal lighting
Temperature and humidity sensing methodTemperature: Class A PT100 armored thermocouple
Humidity: Class A PT100 armored thermocouple
Dry and wet bulb thermometer method (works only in tests with humidity control)
safety protectionFault alarm and cause, processing prompt function, power-off protection function, upper and lower limit temperature protection function, calendar timing function
 (automatic start and automatic stop running), self-diagnosis function
Verify configurationAccess hole with silicone plug (50 mm, 80mm, 100mm left)
Data interfaceEthernet + software, USB data export, 0-40MA signal output

Implementation standards

GB/T5170.2-2008 Temperature test equipment

GB/2423.1-2008 (1EC60068-2-1:2007) Low temperature test method AB.

GB/2423.2-2008(1EC60068-2-2:2007) High temperature test method BA. GB/T5170.5-2008 Damp heat test equipment.

GJB150.3(MIL-STD-810D) High temperature test method.

GJB150.4(MIL-STD-810D) Low temperature test method.

GB2423.3-93 (IEC68-2-3) Test Ca: Constant damp heat test method.

GB2423.4-93(1EC68-2-30) Test Db: alternating damp heat test method.


Equipment security protection

Cooling System

  • Compressor overpressure protection

  • Compressor motor overheat protection

  • Compressor motor overcurrent protection

humidification system

  • Humidification tube dry burning protection

  • Water supply abnormality protection

  • Water shortage abnormality protection

Test chamber protection

  • Adjustable test space over-temperature protection

  • Air conditioning channel extreme over-temperature protection

  • Ultra-high temperature protection is set inside the controller

  • Fan motor overheat protection

Other protection

  • Total power phase sequence and phase loss protection

  • Whole machine leakage protection

  • Load short circuit protection

  • Closed lightning protection


Refrigeration process

Refrigeration Compressors

French Taikang fully enclosed compressor

Way of working

A/B: Mechanical single-stage compression refrigeration system

C: Cascade compressor refrigeration method

Evaporator

High-efficiency components use slope-type evaporators (AC&R compound spoiler-type aluminum fins)

condenser

Adopt water cooling system to facilitate equipment movement, etc.

heat exchanger

Ultra-high efficiency SWEP plate type cold coal cold and heat exchange design is more efficient than the traditional internal spiral type

Installation process

1. Adopt nitrogen flushing welding process, use oxygen-free copper pipes for low-temperature connecting pipes, and use 48-hour high pressure-maintaining and anti-leakage process to ensure welding quality.

2. The pipeline installation adopts the bending process of pipe bending machine. Elbow welding is not allowed in the turning parts.

3. The low-temperature pipeline is equipped with shock-absorbing process pipes, and the refrigeration unit uses shock-absorbing springs and foot pads for shock absorption and noise reduction.

4. The filter dryer absorbs residual moisture, particles, impurities, and acidic substances in the refrigeration system.

5. The selection of solenoid valves is mainly based on brand. Necessary valve core cooling protection measures should be taken when welding solenoid valves to prevent damage or shortening of service life.


FAQ
QIs the price listed on the website the final cost for overseas purchases?
ANo, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
QWhat is the specific process for purchasing the product?
AFirst, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
QWhat should I do if the instrument I purchase needs repair?
AAfter-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.
Knowledge
Standard