| Measurement mode | Automatic range conversion; ordinary measurement, continuous measurement, average measurement |
| Measurement range | 0.00 to 9.99、10.0 to 99.9、100 to 1000 NTU |
| resolution | 0.01 NTU(≤9.99) 0.1NTU(≤99.9) 1NTU(≤1000 |
| Measurement Accuracy | Reading +/- 2% + 0.02 NTU @25 ℃/77 ° F |
| Turbidity Repeatability | +/- 1% of the reading or 0.02 NTU, whichever is greater |
| minimum detection limit | <0.02 NTI |
| Method/standard | Turbidity (90 °) or scattering turbidity (90 ° & 180 °) conforms to USEPA Method 180.1 and Standard 2130B |
| Calibrated mode | Manual calibration, built-in 4-point calibrated points: < 0.1, 15, 100, 750NTU |
| data management | USB-RS232 Dual Interface, 200 sets of measurement Data storage |
| Power mode | 4 × 1.5VAA battery; 15 minutes without any operation, it will automatically shut down. |
| Adapt to the environment | 0°C to 50°C;max 95% RH 无冷凝 |
| Size Weight | Host size: 224 × 87 × 77mm, Host Weight: 512g |
Warranty days: 365 days