Italy

HANNA HI5522 Professional desktop multiparameter Water Quality Analyzer (8 parameters)

SE
Measurement range: -2.000 to 20.000 pH, ORP/mV: +/-2000.0 mV, ISE: 1.00 x 10-6 to 9.99 x 1010 Ion concentration units, linear and non-linear temperature compensation, choice of three salinity measurement ranges, up to four point EC calibrated and single point salinity calibrated.
$2143.00/set
PromotionLoading...
Coupon
Model:

HANNA HI5522 Desktop multiparameter DetectorSPEC

HANNA HI5522 Desktop multiparameter DetectorDetails

HANNA HI5522 Desktop multiparameter DetectorPacking list

SKU
NB045296
Measurement range -pH
-2.0 to 20.0 pH;-2.0 to 20.00;-2.000 to 20.000 pH
Resolution-pH
0.1 pH ;0.01 pH;0.001 pH
Measurement Accuracy -pH
±0.1 pH;±0.01 pH;±0.002 pH±1 LSD
Calibrated mode -pH
Up to five-point acidity pH automatically recognizes the calibrate and five user-defined calibrates with eight built-in acidity standard calibrate points (1.68, 3.00, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45 pH)
Temperature compensation -pH
Automatic or manual temperature compensation, -20.0 to 120.0 ℃, -4.0 to 248.0 ° F, 253.15 to 393.15K
Measurement range -ORP
±2000.0 mV
Resolution - ORP
0.1 mV
Measurement Accuracy - ORP
±0.2 mV ±1 LSD
Measurement range -ISE
1.00X10-6 to 9.99x1010 Ion concentration units
Resolution-ISE
0.001, 0.01, 0.1, 1 Ion concentration unit
Measurement Accuracy - ISE
Readings +/- 0.5% (price ions), readings +/- 1% (divalent ions)
Calibrated mode - ISE
Five automatically recognized calibrates and five user-defined calibrates with five standard calibrated points built in
Temperature Measurement Range
-20.0 to 120.0 ℃、-4.0 to 248.0 ℉、253.15 to 393.15K
temperature resolution
0.1℃、0.1℉、0.1K
Temperature measurement Accuracy
+/- 0.2 ℃、 +/- 0.4 ℉、 +/- 0.2K (ignoring probe error)
EC range
0.000 to 9.999 μS/cm;10.00 to 99.99μS/cm;100.0 to 999.9μS/cm;1.000 to 9.999 mS/cm;10.00 to 99.99 mS/cm;100.0 to 1000.0 mS/cm 自动量程转换
Resolution-EC
0.001 μS/cm;0.01 μS/cm;0.1 μS/cm;0.001 mS/cm;0.01mS/cm;0.1 mS/cm
Measurement Accuracy -EC
Reading +/- 1 %( +/- 0.01 μS/cm)
Calibrated mode - EC
Single point calibrated
Conductivity Detector constant -EC
0.0500 to 200.00/cm
Calibrated mode - EC
Automatic identification of calibrated points, user single-point/multi-point custom calibrated; calibrated expiration reminder;
Temperature compensation -EC
Coefficients 0.00 to 10.00%/℃; linear, non-linear and uncompensated modes
Miscellaneous Indicators EC
Up to 10 set files, 5/channel; USP compliant
TDS range
0.000 to 9.999、10.00 to 99.99、100.0 to 999.9 mg/L(ppm);1.000 to 9.999、10.00 to 99.99、100.0 to 400.0 g/L(ppt);TDS转换系数:1.00
Resolution-TDS
0.001、0.01、0.1 mg/L(ppm)、0.001、0.01、0.1 g/L(ppt)
Measurement Accuracy - TDS
±1 %(±0.01 mg/L)
Measurement range - Electrical resistivity
1.0 to 99.9、 100 to 999 Ohms ·cm;1.00 to 9.99、10.0 to 99.9、100 to 999 KOhms·cm;1.00 to 9.99、10.0 to 100.0 MOhms·cm;自动量程转换
Resolution - Electrical resistivity
0.1、1.0 Ohms·cm;0.01、0.1、1KOhms ·cm;0.01、0.1 MOhms·cm
Measurement Accuracy - Electrical resistivity
+/- 2 %( +/- 1 Ohms · cm), whichever is the largest, ignoring the probe error
Salinity range
Actual salinity: 0.00 to 42.00 PSU; Seawater salinity: 0.00 to 80.00 ppt (g/L); salinity: 0.0 to 400.0% NaCI
Salinity resolution
0.1%, 0.01 ppt (actual salinity, seawater salinity)
Salinity measurement Accuracy
Reading +/- 1%
Salinity calibrated mode
Single point mode (HI7037 salinity 100% calibrated solution can be used)
input channel
1 pH/ORP/ISE+1EC
GLP function
Conductivity Detector constants, reference temperatures, temperature compensation coefficients, calibrating point settings, time, date, and more
Data storage
USB interface, manual storage of 100 groups, 50,000 data/group; 100,000 data/channel; storage type: automatic, manual, destination lock; storage interval: 1 second-80 minutes can be set; additional: 200 USP reports; 200 ISE Method of addition reports (HI5522 only)
display mode
Color display (LCD). With on-screen help, user confirmation, etc
Power mode
230AC/12VDC power adapter
Operating Environment
0 to 50℃(32 to 122 ℉),RH max 55%
Size Weight
Host size: 160x231x94 mm, host Weight: 1.2Kg
HI5522
Desktop multiparameter Detector
Measurement range: -2.000 to 20.000 pH, ORP/mV: +/-2000.0 mV, ISE: 1.00 x 10-6 to 9.99 x 1010 Ion concentration units, linear and non-linear temperature compensation, choice of three salinity measurement ranges, up to four point EC calibrated and single point salinity calibrated.
HANNA HI5522 Professional desktop multiparameter Water Quality Analyzer (8 parameters) Picture

Features

  • Capacitive touch buttons, clear user interface, on-screen operation guide, multiple language options;

  • 5-point pH automatic recognition calibration, 8 built-in standard points, 5 custom calibrations;

  • USP-<645> three-step conductivity analysis standard, in line with international measurement specifications; especially suitable for pure water and ultrapure water measurement;

  • Linear and nonlinear temperature compensation, three salinity measurement range options;

  • Up to four-point EC calibration and single-point salinity calibration;

  • GLP management, USB data interface, charts and curves, convenient data analysis;

  • Data storage, three data storage modes: automatic, manual and end-point lock. Automatically and manually store 100 groups, 50,000 data/group; up to 200 USP data reports;

  • Custom setting files can store or view up to 10 custom files, that is, user-defined settings include: mode, standard point, temperature unit, resolution and other information;

  • User-friendly operation interface supports multiple language selections, and on-screen measurement and calibration operation prompts are convenient and fast.

FAQ
QIs the price listed on the website the final cost for overseas purchases?
ANo, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
QWhat is the specific process for purchasing the product?
AFirst, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
QWhat should I do if the instrument I purchase needs repair?
AAfter-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.
Standard