- SKU
- NB045318
- EC Range 2
- 3.00 to 29.99 mS/cm、30.0 to 200.0 mS/cm、up to 500.0 mS/cm(关闭温度补偿)
- Measurement range of total solids solubility 1
- 0.00 to 14.99 mg/L(ppm)15.0 to 149.9 mg/L(ppm)、150 to 1499 mg/L(ppm)
- Measurement range of total solids solubility 2
- 1.50 to 14.99 g/L, 15.0 to 100.0g/L, up to 400.0g/L (off temperature compensation, TDS factor: 0.80
- Total solid solubility resolution
- 0.01mg/L(ppm)、0.1mg/L(ppm)、1mg/L、0.01g/L(ppt)、0.1g/L(ppt)
- Total solid solubility Accuracy
- Reading +/- 1 %( +/- 0.03 ppm or 1 word, whichever is greater) @25 ℃/77 ° F
- TDS conversion factor
- Factor adjustable range: 0.40 to 0.80, factory default value: 0.50
- Total solid solubility calibration method
- By conductivity calibrated
- Salinity range
- 0.0 to 400.0%NaCl;2.00 to 42.00 PSU ;0.0 to 80.0g/L
- Salinity resolution
- 0.1%NaCl;0.01 PSU;0.01g/L
- Salinity Accuracy
- Reading +/- 1% @25 ℃/77 ° F
- Salinity calibration method
- PSU and g/L (ppt) by conductivity calibrated,% NaCl single point 100% calibrated (recommended HI7037 seawater standard)
- Data storage
- Up to 1000 groups (basic mode: 400 groups) Data storage, manual point storage 200 groups Data storage, manual continuous 200 groups Data storage, time interval (up to 600 automatically, manual 100 samples storage)
- Electrode configuration
- HI763100 built-in temperature sensor four-ring conductivity electrode, 3.5mm Connector, 1 meter wire length
- Measurement range
- -20.0 to 120.0℃( -4.0 to 248.0℉)
- Resolution & Accuracy
- Resolution: 0.1 ℃, 0.1 ° F; Accuracy : +/- 0.5 ℃, +/- 0.9 ° F
- Interface
- Standard USB interface, suitable for fast storage flash drive, Micro USB interface, suitable for data cable transmission charging
- Operating Environment
- 0 to 50 ℃ (32 to 122 ° F), Max95% RH, non-condensing
- Power mode
- 230AC/5VDC power adapter
HI2030
Plate EC-TDS- salinity - temperature meter
Configure HI763100 conductivity dedicated electrode, support manual, manual continuous and time interval three storage modes, with automatic temperature compensation.
Features
Configure HI763100 conductivity special electrode
Measuring range: 0.00 to 500.0mS/cm (actual conductivity) Accuracy: ±1% of reading (±0.05μS/cm) or 1 word
Automatically identify 5 standard solution calibrations and conductivity constant corrections
Three data storage modes: manual, manual continuous and time interval
Automatic/manual range selection and EC/TDS/NaCl reading mode
ATC automatic temperature compensation or off compensation (actual conductivity) GLP records calibration date, time, conductivity constant, etc.
EC and TDS conversion factors with temperature correction available
Three seawater salinity unit options: %NaCl, PSU and g/L
Packing list:machine X1, manual X1, certificate X1, warranty card X1
[Note]Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.