Temperature range High temperature: room temperature~ + 500 ℃; Low temperature: -60 ℃~ room temperature (optional), deformation Measurement range 5mm, deformation measurement resolution 0.001mm, high technical content, strong software function, Measurement range, accurate measurement, stable performance, easy to operate.
High temperature: room temperature~ + 500 ℃; low temperature: -60 ℃~ room temperature (optional)
Measurement range
5mm
Deformation measurement resolution
0.001Mm (Displacement Sensors)
Heating rate
0.5 ℃/min, 1.2 ℃/min, 2 ℃/min, 5.0 ℃/min (can be set arbitrarily between 0.5-5 ℃/min)
Cooling rate
-0.5 ℃/min~ -2 ℃/min. (with refrigeration)
Heating rate error
≤0.5℃/min
Temperature measurement Accuracy
±0.5℃
Temperature control Accuracy
±1℃
load range
50g-3000g
Required space
front and rear 0.5m, left and right 1.0m
display function
Real-time display of temperature~ deformation and temperature~ time Linear dispersion
computing function
Automatic calculation of Tg, Tf and expansion coefficient
Print function
Printing test Linear dispersion and test report
Instrument size
438mm×438mm×680mm
Power
220V±10%, 50HZ, 800W
XWJ-500B
Thermomechanical Analyzer
Temperature range High temperature: room temperature~ + 500 ℃; Low temperature: -60 ℃~ room temperature (optional), deformation Measurement range 5mm, deformation measurement resolution 0.001mm, high technical content, strong software function, Measurement range, accurate measurement, stable performance, easy to operate.
Introduction
The XWJ-500B thermomechanical analyzer has a wide measurement range, diverse measurement methods, powerful analysis functions, and accurate and reliable data. It is a very reliable analytical instrument suitable for product development. It is suitable for plastic production enterprises to conduct product quality analysis and adjust product formulas, etc. It can also be used by plastic formula R&D institutions to adjust formula analysis.
This instrument is an analytical instrument used to measure the glass transition temperature (Tg) and flow temperature (Tf) of polymers. This product heats the sample at a certain heating rate, causing the sample to deform as the temperature increases under a constant, small load, and measures the temperature-deformation curve of the sample. By analyzing the temperature-deformation curve of the sample, the sample is measured. Such glass transition temperature (Tg) and flow temperature (Tf). It has the advantages of high technical content, strong software functions, wide measurement range, accurate measurement, stable performance and easy operation.
fit standard
GB11998-89 "Method for determination of glass transition temperature of plastics - Thermomechanical analysis method"
Packing list:
[Note]Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
QOur Plastics Factory wants to measure the Glass Temperature of our products. Which Thermal Mechanical Analyzer should we choose?
A
The XWJ-500B thermomechanical analyzer is specially designed for the plastics industry. It can accurately measure the Glass transition temperature Tg and flow temperature Tf to help you optimize the formula and improve the product mass. The instrument conforms to the national standard GB11998-89, with simple operation and reliable data. It is suitable for daily inspection and R & D requirements of plastic manufacturers.
QThe laboratory needs to test a variety of deformation modes. Is there a suitable instrument recommended?
A
The XWJ-500B Thermomechanical Analyzer supports five measurement forms of compression, bending, stretching, penetration and expansion deformation, covering common polymer testing requirements. Temperature range from -60 ° C to 500 ° C (low temperature optional), deformation Measurement accuracy is high, can flexibly adapt to the analysis of different materials.