- SKU
- NB004657
- Impact energy
- 90mJ
- Impact device Mass
- 20.0g
- Ball Hardness
- 1600HV
- Ball head diameter
- 5mm
- Ball head material
- tungsten carbide
- Impact device diameter
- 30mm
- Impact device length
- 254mm
- Impact device Mass
- 250g
- Maximum Specimen Hardness
- 650HB
- Specimen Surface Roughness
- 6.3µm
- Minimum specimen Weight
- >15kg
- Need stable support
- 5~15kg
- Dense coupling is required
- 0.5~5kg
- Minimum test Film thickness
- 10mm
- Minimum depth of hardened layer
- ≥1.2mm
- Hardness 300HV Indentation Diameter
- 1.03mm
- Hardness 300HV Indentation
- 53µm
- Hardness 600HV Indentation Diameter
- 0.90mm
- Hardness 600HV Indentation
- 41µm
- Application of impact device
- G-type measurement of large, heavy and rough surface castings and forgings
G type
Impact device (Hardness Gauge)
G type impact device introduction
Leeb G-type Hardness Tester impact device The impact device is an optional accessory for Leeb120 and Leeb140 Leeb Hardness Testers, and is used to measure the hardness of gray cast iron/nodular cast iron.
Product testing range
| Material | Hardness system | Impact device |
| D/DC | C | G | DL |
| steel and cast steel | HRC | 17.1~68.5 | 20.0~69.5 |
| 20.6~68.2 |
| HRB | 59.6~99.6 |
| 47.7~99.9 | 37.0~99.9 |
| HB | 140~651 | 80~683 | 90~646 | 150~646 |
| HV | 83~976 | 80~996 |
| 85~950 |
| HS | 26.4~99.5 | 31.9~102.3 |
| 26.5~99.4 |
| forged steel | HB | 142~651 |
|
|
|
| alloy tool steel | HRC | 17.1~67.1 | 20.0~69.5 |
|
|
| HV | 83~976 | 80~996 |
|
|
| Stainless steel | HRB | 59.6~99.6 |
|
|
|
| HB | 140~651 |
|
|
|
| HV | 83~976 |
|
|
|
| grey cast iron |
| 140~334 |
| 92~326 |
|
| Ductile Iron | HB | 140~387 |
| 140~364 |
|
| cast aluminum alloy | HB | 30~159 |
|
|
|
| copper zinc alloy | HB | 40~173 |
|
|
|
| Copper-tin alloy (bronze) | HB | 60~290 |
|
|
|
| pure copper | HB | 45~315 |
|
|
|
Pictures


Packing list:Impact device x1, Certificate x1, Manual x1, Encasement box x1