China

HUXI HLX-4009G high-low temperature Refrigerated Circulating Chiller 9L/-40~100 ℃

SE
Water tank volume 9L, Temperature range -40~100 ℃, Stability +/- 0.5 ℃, Cooling capacity 1500W, Pump Flow rate 22L/min, Pump pressure 1bar
$1930.00/set
PromotionLoading...
Coupon
Model:
Certificate

HuXi HLX-4009G High-low temperature coolant Circulating ChillerSPEC

HuXi HLX-4009G High-low temperature coolant Circulating ChillerDetails

HuXi HLX-4009G High-low temperature coolant Circulating ChillerPacking list

SKU
NB046627
Water tank Volume
9L
Temperature range
-40~100℃
Temperature Fluctuation value
±0.1℃
Set Accuracy
±0.01℃
Display Accuracy
±0.001
Refrigeration power
1200W
cooling Capacity
1500W
refrigerant
R404A
Heating power
2000W
Display temperature mode
TFT
setting method
touch screen
High temperature drop
support
safety protection
Low liquid level protection, high temperature protection, over temperature protection
Environmental temperature
5~32℃
Power supply Voltage
200-230V
Circular mode
External circulation
Flow rate (max/rated)
22L/min
Loop Interface
10Mm pagoda Connector
Net Weight
35kg
Gross Weight
43kg
Dimensions (W * D * H)
300*500*620mm
Outer box size (W * D * H)
380*580*800mm
HLX-4009G
High-low temperature coolant Circulating Chiller
Water tank volume 9L, Temperature range -40~100 ℃, Stability +/- 0.5 ℃, Cooling capacity 1500W, Pump Flow rate 22L/min, Pump pressure 1bar
HUXI HLX-4009G high-low temperature Refrigerated Circulating Chiller 9L/-40~100 ℃ Picture

Introduction

HLX series high-low temperature cooling Circulating Chiller is a circulation device that provides heat and cold sources for reactors, rotary steaming, etc., and has laboratory instruments and equipment with dual functions of heating and refrigeration. It is mainly used in the fields of chemical, pharmaceutical and biological fields supporting glass reactors, rotary evaporators, fermentation tanks, calorimeters, and is used in petroleum, metallurgy, medicine, biochemistry, physical properties, testing and chemical synthesis and other research departments, colleges and universities, factory laboratories and metrology and quality inspection departments.

Features

  • The imported compressor is equipped with proportional refrigeration technology and PID technology, and the refrigeration system has a long service life and is more energy-saving;

  • Built-in pressure suction pump, large flow, long head, stable water flow, can withstand the temperature of 80~300 °C for a long time;

  • High temperature direct drop technology (CIT technology), which can directly cool down even at high temperatures;

  • The pump head supports short-term idling to prevent operation errors and can test the equipment;

  • Equipped with a liquid drain port to prevent accidents;

  • Resistant to metal and non-metal hard impurities and fibers brought in from the outside, it has a long service life;

  • Imported temperature sensor: four-wire imported PT100 temperature sensor, which can be calibrated at single-point, two-point and three-point temperature;

  • Low liquid level protection: light and sound can be reminded at the same time when the alarm is made.

FAQ
QThe laboratory needs a high and low temperature circulation equipment for the Rotary evaporator. Do you have any recommendations?
A HUXI HLX-4009G high and low temperature coolant circulating pump is recommended. Specially designed for rotary steaming, Temperature range -40~ 100 ℃, temperature control accuracy +/- 0.1 ℃, can quickly switch heating and cooling. Built-in large Flow rate pump ensures evaporation efficiency, high temperature direct drop technology allows you to directly cool down from high temperature without waiting, suitable for solvent recovery process of rotary steaming.
QReaction Kettle requires precise temperature control during chemical synthesis. Which cycle equipment should I choose?
A HUXI HLX-4009G is a reliable choice. It uses PID technology and imported temperature sensor, temperature control accuracy of +/- 0.01 ℃, temperature fluctuation is only +/- 0.1 ℃, can meet the strict temperature requirements of synthesis reaction. The pump head is resistant to high temperature up to 300 ℃, supports short-term idling, avoids damage to equipment due to operation errors, and ensures the continuity and safety of synthesis experiments.
QCan this equipment meet both heating and cooling needs? In which scenarios is it suitable?
A Yes! HLX-4009G is a heating and cooling dual-function all-in-one machine, covering common experimental needs in a wide temperature range of -40~ 100 ° C. It is suitable for temperature control of Reaction Kettle, rotary steaming, fermentation tank and other equipment, and can be used in chemical, biological, petroleum and other fields. Its high temperature direct drop technology and low liquid level protection function are suitable for experimental scenarios that require rapid temperature change or long-term operation.