| Display mode | 5 inch TFT true color LCD industrial touch screen 800 * 480 pixels |
| display format | Real-time list, Linear dispersion (historical Linear dispersion), bar chart, current alarm (historical alarm), file record list. |
| record query | This machine view analysis records Linear dispersion, historical alarm records, computer software analysis query. |
| Thermocouple | K JETNSRB |
| Basic Precision | 0.2 ℃ + 2 characters (without Thermocouple error) |
| Measurement range | -200~1820C (subject to Thermocouple indexing range) |
| cold junction compensation | Accuracy: 0.5 ℃ |
| resolution | 0.01℃ |
| correction | Independent error correction per channel Y = kx + b (x = Measured value) |
| TYPE-C interface | 0TG function, integrating charging, PC communication, connecting U disk, and integrating U disk |
| Number of files | 100 (cyclic record) |
| file Capacity | A file can record the number 120,000 groups (without distinguishing Number of channels), a total of 12 million groups. |
| recording interval | 100mS-2h: 59m: 59S.950mS Any setting (minimum 0.1 second when opening 2 channels) |
| recording time | 0.1 second interval, 13.88 hours of continuous recording, total duration = recording interval X138.8 hours. |
| Sampling speed, filtering | Single channel fastest: 50mS, filter coefficient (1-64) larger Measurement speed slower, more stable value |
| Communication interface | Wireless W1FI (2.4G), TYPE-C (USB) standard analysis software |
| channel isolation | AC/DC up to 350V, high Voltage live measurement, super anti-interference ability |
| Alarm sound | Buzzer sound all the way (sound at any alarm, can be set to mute) |
| Communication interface | Wireless WIFI, TYPE-C (USB) |
| Power supply | Built-in 5000mAH battery, 10 hours (turn off WIFI), with charger, support 18W charging. |
| Thermocouple | Standard per channel - 2 m K type Thermocouple |
| size | 1 95X127X41mm |
| Weight | About 1.5Kg |
| environmental conditions | 5~ 40 ℃, 20%~ 80% RH (no condensation) |
Warranty days: 730 days
1. For product quality problems caused by non-human products of instruments and equipment, the manufacturer will provide free warranty services during the warranty period;
2. There is no warranty period for fragile and consumable parts, and if you find product quality problems after receiving the goods, please apply for return and exchange within 7 days, and you will not wait for the overdue ones;