| Measurement parameter | COD, ammonia nitrogen, total phosphorus, total nitrogen, turbidity, suspended solids |
| measurement error | ≤± 5% |
| Sexual repeability | ≤±5% |
| Optical inspection stability | ≤ +/- 0.001A/20min (100,000 hours life) |
| data storage | 30,000 records (including item, date, time, sample ID, etc.) |
| Linear dispersion quantity | 96 Linear dispersion (expandable) |
| printer | Built-in thermal printer |
| Display | 5-Inch high-definition color LCD screen |
| batch Volume | 16/25 water samples/batch |
| Colorimetric method | Digestion Colorimetric tube |
| sample compatibility | 16 mm (round), 25mm (round) |
| communication interface | USB cable, LAN, etc |
| exception notification | Blank comparison prompt, over-range prompt, etc |
| operation interface | Membrane push-button operation, Chinese/English (optional) |
| Power Supply | AC(220V±10%),50Hz |
| power | 5W |
| protection class | IP30 |
| Environment humidity | ≤ 85% non-condensing |
| Environment temperature | 5~40℃ |
| Dimensions | 348*250*145mm |
| instrument Weight | Main unit less than 2.6kg |
| COD detection method | Method 1: HJ/T399-2007 standard; Method 2: Import detection method |
| COD Measurement Range | 0-20000mg/L;0-20000mg/L |
| Lower limit of COD detection | 3 mg/L;2 mg/L |
| COD digestion environment | 165℃,15min;165℃,20min |
| COD anti-chlorine interference | 1000 mg/L (expandable); 1000 mg/L (expandable) |
| Ammonia nitrogen detection method | HJ535-2009 "Water quality, determination of ammonia nitrogen, Nessler reagent Spectrophotometric method, spec-trophotometry" |
| Ammonia nitrogen Measurement Range | 0-50 mg/L |
| Ammonia nitrogen detection lower limit | 0.01mg/L |
| Ammonia nitrogen digestion environment | No need to digest |
| Total phosphorus detection method | GB11893-89 |
| Total phosphorus Measurement Range | 0~20mg/L |
| Total phosphorus detection limit | 0.01mg/L |
| Total Phosphorus Digestion | 125℃,30min |
| Turbidity detection method | Spectrophotometric method,spec-trophotometry |
| Turbidity Measurement Range | 0~1000NTU |
| Lower limit of turbidity detection | 0.5NTU |
| Turbidity digestion environment | No need to digest |
| Suspension detection method | Direct Colorimetry |
| Suspension Measurement Range | 0~500mg/L |
| Lower limit of suspended solids detection | 2mg/L |
| Suspension digestion environment | No need to digest |
| TN detection method | Persulfate oxidation |
| TN Measurement Range | 0~250mg/L |
| TN detection lower limit | 0.5 mg/L |
| TN digestion environment | 125℃,30min |
None