China

CHIJIU H01-2 Digital Magnetic Stirrer Stir Capacity 50L Digital Display Speed

SE
Max. stirring Capacity 50L, stirring speed 60~1400rpm, input power 100W
$501.00/set
PromotionLoading...
Coupon
Model:

ChiJiu H01-2 Magnetic StirrerSPEC

ChiJiu H01-2 Magnetic StirrerDetails

ChiJiu H01-2 Magnetic StirrerPacking list

SKU
NB005419
Max. Capacity of stirring
50L
Stirring speed
60~1400rpm
Input power
100W
Dimensions
410*400*115mm
Power
AC220V±10% 50HZ
Operating Mode
continuous
H01-2
Magnetic Stirrer
Max. stirring Capacity 50L, stirring speed 60~1400rpm, input power 100W
CHIJIU H01-2 Digital Magnetic Stirrer Stir Capacity 50L Digital Display Speed Picture

Product Description

The H01-2 digital display magnetic stirrer is a large-capacity, high-power single-stirring magnetic stirrer, which has a solid structure and is driven by a three-phase brushless DC motor with large torque and strong magnetic force. It is also equipped with a digital display speed, suitable for large volumetric stirring.

product display

Chijiu H01-2 Digital Display Magnetic Stirrer Display

Instructions

1. Test run: place the H01-2 digital display magnetic stirrer stably, connect the power cord, plug in the power supply, turn on the power switch, the built-in indicator light of the switch is on, and the digital display is "0000", adjust the knob clockwise, and the digital display shows the speed , after the test run is normal, turn the speed to the lowest and shut down.

2. When using, put the container in the middle of the stirrer, and put in the stirrer, so that the stirrer is in the center, then turn on the machine, adjust the knob, select the required stirring speed, and gradually accelerate or decelerate when adjusting the speed. Note: The container needs to be made of non-ferromagnetic material with a smooth bottom and a thickness not greater than 10mm.

3. If the stirring bar is lost during use, it should be turned off immediately to prevent the stirring bar from demagnetizing.

Precautions

1. The power supply should be well grounded.

2. The central Φ110mm of the working surface of the agitator should not be able to withstand large pressure.

3. The liquid medicine in the container should not be overfilled, so as not to overflow when stirring.

4. It is forbidden to put the stirrer directly on the stirrer for trial operation.

5. Due to the strong magnetic force of the large-capacity magnetic stirrer, the friction between the stirrer and the bottom of the container is relatively large. The user should use a container with a small friction coefficient as much as possible, and always pay attention to the wear of the stirrer. If the stirrer is severely worn, The friction will increase, overloading the motor or even stalling and damaging the motor. Therefore, you should always observe the display status of the tachometer during use. If you find that the speed drops, suddenly stops, or it is difficult to start, you should shut down the machine immediately.

6. The higher the rotation speed in use, the faster the agitator will wear out, so the rotation speed should not be too high, as long as the stirring effect can be achieved.

FAQ
QThe stirring liquid is about 50L, and the stirring speed reaches 1000rpm. Is this satisfactory?
A Yes, ha.
QCan this model of Magnetic Stirrer be used for stirring solutions in slightly raised glass bottles?
A This Stirrer requires that the bottom needs to be flat, and the bulge cannot be stirred. As long as the stirred liquid has flowability, it can be stirred.
QMeet GB/T33546 Magnetic Stirrer
A H01-2 Magnetic Stirrer, please refer to it.
Knowledge
Standard