JinKeLian STM-III Stormer viscometer is designed in accordance with ASTM standards and GB/T9269-88 standards, using a single-chip microcomputer, which can directly display the test results through the display screen, and the operator can directly read the test samples without looking up the table KU value. The volume of STM-III Stormer viscometer is about 500ml, the measurement range is 53~141KU, the measurement is stable and accurate, and the operation is simple.
STM-III Stormer viscometer is suitable for measuring the viscosity of architectural coatings, water-based coatings, inks, paints and other products expressed by KU value.
1. Measurement environment: 23℃±0.2℃
2. Put the STM-III Stormer viscometer from the box on the side of the workbench, so that the weight bracket can slide down to the ground smoothly.
3. Plug in the power plug, and click the power switch on the lower right side of the back cover. The number on the four-digit display will show zero. At the same time, the speed indicator is on, and the STM-III Stormer viscometer is in the working state of detecting the speed.
4. Fully stir the sample to be tested and put it into the container, so that the liquid level of the container mouth is about 19mm, then put it on the container support, immerse the rotor in the liquid, so that the liquid level just reaches the mark on the rotor shaft. Then tighten the locking screw.
5. Put the weight on the STM-III Stormer viscometer weight bracket, loosen the brake handle, the rotor will start to rotate, and at the same time, the counter will start to display the rotor speed per minute on the panel. If the rotating speed is lower than 200 rpm, the weight of the weight should be increased, and if it exceeds, the weight of the weight should be reduced. Repeat this step several times to stabilize the rotor speed at 200 rpm. At this time, the total weight of the weight (including the bracket 30g) is a constant load.
6. Press the number key and the shift key, and input the total amount of weight (g) added at 200 rpm consistent with the displayed number. Then press the replacement key, at this time the indicator light of the KU value is on, and the number displayed on the display is the KU value of the sample to be tested. (The shift indicator point cannot be mistaken for a decimal point).
7. Press reset to turn on the speed indicator light, and the STM-III Stormer viscometer is in the state of detecting the speed again.
Immerse the paddle rotor into the sample to be tested, increase or decrease the weight of the weight to keep the speed of the paddle rotor at 200 rpm, and input the weight of the sample to be tested at 200 rpm and replace it with viscose by a single chip microcomputer. Consistency KU value.

Remarks: For the operation instructions of STM-III Stormer Viscometer, please click "STM-III Stormer Viscometer Instruction Manual"