China

MODERNER ZW-I Conical Mandrel Bending Tester - Elongation test of adhesive coatings for paints, varnishes, etc

SE
GB11185-89 Test to evaluate the performance of coatings of Paints and varnishes against cracking or disbonding from a metal baseplate when bent about a tapered shaft under standard conditions.
$487.60/set
PromotionLoading...
Coupon
Model:
Certificate

Moderner ZW-I Conical Mandrel Bending TesterSPEC

Moderner ZW-I Conical Mandrel Bending TesterDetails

Moderner ZW-I Conical Mandrel Bending TesterPacking list

SKU
NB000690
Tapered shaft size
F3.2 ~ 38mm
Overall dimensions
280*110*120mm
Weight
9kg
ZW-I
Conical Mandrel Bending Tester
GB11185-89 Test to evaluate the performance of coatings of Paints and varnishes against cracking or disbonding from a metal baseplate when bent about a tapered shaft under standard conditions.
MODERNER ZW-I Conical Mandrel Bending Tester - Elongation test of adhesive coatings for paints, varnishes, etc Picture

Overview of ZW-I Cone Bending Tester

ZW-Ⅰ Cone Bending Tester is used to evaluate the cracking resistance or peeling performance of the paint and varnish coatings when they are bent around a cone shaft under standard conditions according to the national standard GB11185-89. It can also refer to the standard ASTM D522-60 (1979) of the American Society for Testing and Materials for the determination of the elongation test of the adhesive coating coated on the flat metal with a uniform surface structure.

real shot

Real shot of ZW-I cone bending tester in MODERNER Figure 1

Real shot of ZW-I cone bending tester in MODERNER Figure 2

Real shot of MODERNER ZW-I cone bending tester Figure 3

Real shot of MODERNER ZW-I cone bending tester Figure 4

Specifications

Cone size (GB11185-89 regulation): large end Φ38, small end Φ3.2, length 203mm

Dimensions (without handle): 280x110x120mm

Weight: 9kg

operation method

1) Test plate specification: according to the national standard GB11185-89, it is 100mm×180mm, with a maximum thickness of 0.8mm, polished steel, polished tin plate or soft aluminum plate. According to the American Society for Testing and Materials standard ASTM D522-60 (1979), it is a cold-rolled carbon steel with a thickness of 114mm×177mm and a maximum thickness of 0.8mm. The edges of the panels bent in the test were lightly rounded to remove burrs before the topcoat was applied. The design of the ZW-I conical bending tester considers the metal plate with a maximum thickness of 0.8mm. Do not use a test plate that exceeds the standard, so as not to damage the instrument.

2) Test environment and times: Unless otherwise agreed, the test shall be carried out under the conditions of 23°C±2°C and relative humidity of 50±5%. At least three identical samples shall be tested.

3) Bending test steps: Turn the handle of the instrument to the front (where the scale is attached); turn the unpainted surface of the test plate toward the cone shaft, pass through the gap between the rotating shaft and the cone shaft, and insert it into the seam of the pressure plate; place the test plate One short side of the cone lightly rests on the step surface of the small end of the cone shaft; lock the wing nut to press the test plate tightly; turn the handle evenly and smoothly to bend the test plate so that it can rotate 180° around the shaft within 2 to 3 seconds; loosen Unscrew the wing nut and remove the test panel.

4) Check the test plate according to the standard. The test plate is a user-owned spare part.

maintenance

After use, wipe the instrument clean, unscrew the handle, and put it into the wooden box of the instrument.

FAQ
QDo you have a test sheet metal bending instrument? Thickness over 1mm.
A No, the Bending tester is only suitable for samples with a thickness below 0.8mm.
QIs this ZW-I Bending tester compliant?
A Hello, the ZW-I cone Bending tester of modern environment meets the GB11185 standard.
QIs the operation of ZW-I Conical Mandrel Tester complicated?
A Hello, it's not complicated. You can refer to the operation manual on this product page.
QI want to do a test on the ASTMD522 standard. Is this Conical Mandrel Bending Tester okay?
A Sure
QThis ZW-I bending test machine on the experimental plate Film thickness requirements?
A If required, the maximum thickness shall not exceed 0.8mm according to ASTM D522-60
QWhat are the specific requirements of the ZW-I tapered Bending tester for the test plate?
A Hello, according to the national standard GB11185-89, the size of the test plate is 100mm × 180mm, and the maximum Film thickness is 0.8mm polished steel, polished tinplate or soft aluminum plate.
Standard