China

YouKe UV754 UV-Visible Spectrophotometer Wavelength range 200~1000nm

YouKe UV754 UV-Visible Spectrophotometer adopts imported long-life tungsten lamp and deuterium lamp, Light spectrum Bandwidth is 4nm, UV754 adopts enhanced 8-bit processor, and the main frequency reaches 24M.
Model:
Certificate

Yoke UV754 SpectrophotometerSPEC

Yoke UV754 SpectrophotometerDetails

Yoke UV754 SpectrophotometerPacking list

SKU
NB007218
Light spectrum Bandwidth
4nm
Wavelength range
190~1100nm
Wavelength Precision
±2nm
Wavelength Repeatability
0.5nm
stray light
0.3%T@220nm,360nm
Stability
±0.002A/h
monitor
128 * 64 check point array LCD display
Transmittance Precision
±0.5%T
Transmittance Repeatability
0.2%T
Measurement range of transmittance
0.0~200%T
Absorption range
-0.030-3.0A
UV754
Spectrophotometer
YouKe UV754 UV-Visible Spectrophotometer Wavelength range 200~1000nm Picture

Introduction

The UV754 UV-Vis Spectrophotometer can do qualitative and quantitative analysis of sample substances in the ultraviolet and visible spectrum regions, and can be widely used in clinical testing, biochemistry, petrochemical, environmental testing, food and quality control departments, and can be used as teaching demonstration and experimental instruments for related courses in colleges and universities.

Features

1. 128*64 dot matrix large-screen LCD display

2. Adopt enhanced 8-bit processor, the main frequency reaches 24M

3. The optimized CT optical path ensures ultra-low stray light

4. Imported long-life tungsten lamp and deuterium lamp

5. Simple lamp source replacement operation

Features

1) Automatic control function

● Self-check of the working status of the internal system of the instrument when it is turned on

● Wavelength is automatically calibrated

● Automatic calibration of color filters

● The light source is automatically switched

● Displays a variety of error messages

● Data and graph storage and printing

2) Analyze the test function

● Photometry: Measure the absorbance or transmittance of a sample

● Quantitative analysis: A standard curve is established using a standard sample, and the concentration of an unknown sample is measured using this curve, and if the slope and intercept of the curve are known, the coefficient method can also be used

Diagram of the internal structure

Structure diagram of Yoke UV754 UV-Vis Spectrophotometer

Pictures