| Optical inspection system | Double beam of light system |
| Light spectrum bandwidth | 2nm |
| Wavelength range | 190~1100nm |
| Wavelength Accuracy | +/- 0.1Nm (D2 656.1nm ), +/- 0.3nm full band |
| Wavelength Repeatability | ≤0.1 nm |
| Photometric range | 0~200%T、-4~4A |
| Photometric Accuracy | ±0.002A(0~0.5A)、±0.004A(0.5~1A)、±0.2 %T (0~100 %T) |
| Luminosity Repeatability | ±0.001A(0~0.5A);±0.002A(0.5~1A);±0.1%T(0~100%T) |
| stray light | ≤0.03%T |
| Noise level | ≤0.0004 Abs/h |
| Stability | ±0.0003A(500nm 处) |
| Baseline flatness | ≤±0.0005A |
| Operating Mode | T、A、C、E |
| keyboard | Membrane keys |
| Interface | Parallel interface or USB interface |
| display | 320 * 240-Bit high-brightness 6 "large-screen LCD |
| Light source | 190 nm~1100 nm |
| Detector | Imported silicon photodiode |
| Power | AC 220V/50Hz 或 AC 110V/60Hz |
| Weight | 30kg |
| Drift | ≤± 0.0004 Abs/h |
Warranty days: 365 days
1. For product quality problems caused by non-human products of instruments and equipment, the manufacturer will provide free warranty services during the warranty period;
2. There is no warranty period for fragile and consumable parts, and if you find product quality problems after receiving the goods, please apply for return and exchange within 7 days, and you will not wait for the overdue ones;