China

SDC STM-IV Stormer Viscometer Viscosity Range 40~141KU

The SDC STM-IV Stormer Viscometer is a digital Stormer Viscometer with Krebs as the Measurement unit. The STM-IV viscosimeter has a Measurement range of 40~141KU and is equipped with a 500ml Glass container.
Model:

SDC STM-IV Stormer ViscometerSPEC

SDC STM-IV Stormer ViscometerDetails

SDC STM-IV Stormer ViscometerPacking list

SKU
NB008242
Display mode
digital display
Measurement range
40~141KU,32~1099g
Unit display
KU/g
Swirl/spin speed
200rpm
Repeatability
±1%F.S
Precision
±1.5%F.S
Power supply Voltage
220V AC 50Hz 16W
Applicable standards
ASTM D562
Environmental temperature
0~40℃
Dimensions
150*110*750mm
Instrument Weight
11kg
printer output
Print through the host computer software
Container size
500ml
Working Relative Humidity
20~80% non-condensing
Speed Accuracy
±0.1rpm
Lifting method
Lever lift
STM-IV
Stormer Viscometer
SDC STM-IV Stormer Viscometer Viscosity Range 40~141KU Picture

Introduction

The Sedchang STM-IV Stormer viscometer is a digital display Stormer viscometer used to measure the viscosity of fluids in Krebs units.The measured viscosity range is 40~141 KU,The display can be displayedViscosity values in Krebs units (KU), gram units (gm).The STM-IV Stormer viscometer is a 200rpm constant speed motor that drives a paddle rotor to rotateReaction torque at 200RPM is converted to Krebs singleThe viscosity value of the bit.

Product composition:

The STM-IV Viscometer includes the following accessories

(1) STM-IV Viscometer Vertical Rods and Base                                      2One 500ml glass container

(3) Propeller rotor (4Power cord 250 VAC one

(5) 500ml glass container holder small (6) data cable HL-340

(7) The 500ml glass container holder is large                                              (8Operating Manual (SDC/04-559)


Front view and side view of the Sedchang STM-IV Stormer viscometer

Rear view of the Sedchan STM-IV Stormer viscometer

Product display diagram

Detail of the control head of the Sedchang STM-IV Stormer viscometer

Detail of the Sedchan STM-IV Stormer viscometer impeller


FAQ
QHelp me find a Stormer Viscometer that meets the ASTMD 562 standard.
A This STM-IV Stormer Viscometer is fine!
QWhat happens when STM-IV viscosity measurement exceeds the viscosity range?
A The instrument can not measure more than the maximum range of the sample, otherwise it will affect the accuracy of the instrument.
QExcuse me, why is the boot screen locked after my STM-IV Stormer Viscometer is turned off?
A Hello, do you turn it on as soon as you turn it off? If the instrument is turned off and then turned back on immediately (that is, the closing time is less than 10 seconds), the display screen and swirl/spin will be temporarily locked.
QIs the gram scale a unit of viscosity?
A Hello, no, the gram scale represents the weight required to drive the bladed rotor to swirl/spin at 200RPM in the test liquid. The gram scale is not a unit of viscosity.
QCan the STM-IV Stormer Viscometer test the viscosity of the rolling paste?
A Hello, STM-IV can be tested, but it needs to be used with an additional paste rotor, which is suitable for high consistency materials, such as roller rolling paste.
Knowledge