NIRUN DV-79 Digital Viscometer Viscosity Range 1-1 million cP
SE
Digital display, Measurement range 50~1000000mPa · s, rotor specifications E, F, G three rotors, Rotor speed 7.5/75/750rpm, reproducibility +/- 0.5% F.S, Measurement accuracy +/- 1% F.S, Power supply Voltage 100~240V 50~60Hz, Temperature range 0~120 ℃, measurement container B container, temperature sensor RTD temperature sensor, timing function.
Digital display, Measurement range 50~1000000mPa · s, rotor specifications E, F, G three rotors, Rotor speed 7.5/75/750rpm, reproducibility +/- 0.5% F.S, Measurement accuracy +/- 1% F.S, Power supply Voltage 100~240V 50~60Hz, Temperature range 0~120 ℃, measurement container B container, temperature sensor RTD temperature sensor, timing function.
Technical advantages
Stepper motor direct drive, no gear, accurate speed
LCD liquid crystal with blue backlight, displaying a large amount of information
It can display the maximum viscosity value that the rotor can measure at the current speed.
Measurement Timing Function
Viscosity value shows continuous change
There is an alarm sound prompt when the measurement range is exceeded
Standard RTD temperature probe
Computer full range linear calibration
AC power adapter 100V-240V, good anti-interference
Applications
DV-79 digital viscometer is widely used in starch, glue , latex , solvent-based adhesives , polymer solutions , petroleum , paint coatings , solvents , cosmetics , milk products , medicines , fruit juices, etc.
Pictures
Optional
DV data acquisition and program-controlled analysis software
Special viscosity measurement constant temperature Water Tank
[Note]Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
QI am using a pointer now. What is the difference between the digital display and the pointer? Do you also need to circulate water for the digital display?
A
The pointer Viscometer is a traditional model that uses the pointer to display the viscosity value, while the digital display Viscometer uses a digital display to display the viscosity value, which has higher accuracy and convenience. Temperature has a great influence on viscosity measurement, so it is necessary to circulate water to control thermostatic conditions.
QThe viscosity reaches hundreds of thousands. What kind of Viscometer test do you use? It exceeds 500,000 mpas.
A
Recommendation NIRUN SNB-2 and NIRUN DV-79
QWhat is the difference between Nerun DV-79 and NDJ-79 Digital Viscometer?
A
Hello, DV-79 digital Viscometer is an upgrade of NDJ-79 model, and DV-79 is an LCD digital display screen.
QI want to do a test on the ISO 12634 standard. Is this digital Viscometer OK?
A
Sure
QWhat happens if the Viscometer is outside the measurement range?
A
When the current speed and the current rotor measurement range are exceeded, the screen will display and a beeping alarm sound will be issued. When the torque exceeds 100%, the percentage reading and viscosity reading are all EEEE.
QIs the NDJ-8S digital rotational viscometer suitable for measuring industrial wastewater? (Contains various salts, organic matter, garbage permeate, etc.)
A
NDJ-8S can not be tested, because the low speed, the sample is prone to sedimentation. It is recommended that DV-79 digital Viscometer with A, B containers.