NIRUN DV-79A Digital Viscometer Low Viscosity

SE
Display mode: digital display, Measurement range: 30~1500000mPa · s, rotor specifications E, F, G three kinds of rotors, Rotor speed: 5~ 750rpm, reproducibility : +/- 0.5% F. S, Measurement accuracy : +/- 1% F. S, measurement container B container, Temperature range 0~120 ℃.
$1737.00/set
PromotionLoading...
Coupon
Model:
Certificate

NiRun DV-79A Digital ViscometerSPEC

NiRun DV-79A Digital ViscometerDetails

NiRun DV-79A Digital ViscometerPacking list

SKU
NB008768
Display
digital display
Measuring range
30~1500000mPa·s
Rotor Specifications
E, F, G three rotors
Rotor speed
5~750rpm
Reproducibility
±0.5%F.S
Measurement Accuracy
±1%
Power
110~220V 50Hz/60Hz
Measurement container
B container
Temperature range
0~120℃
temperature probe
RTD temperature probe
Timer
Yes.
Speed control method
In 100rpm increments
DV-79A
Digital Viscometer
Display mode: digital display, Measurement range: 30~1500000mPa · s, rotor specifications E, F, G three kinds of rotors, Rotor speed: 5~ 750rpm, reproducibility : +/- 0.5% F. S, Measurement accuracy : +/- 1% F. S, measurement container B container, Temperature range 0~120 ℃.
NIRUN DV-79A Digital Viscometer Low Viscosity Picture

Technical advantages

  • Stepper motor direct drive, no gear, accurate speed

  • LCD liquid crystal with blue backlight, displaying a large amount of information

  • It can display the maximum viscosity value that the rotor can measure at the current speed.

  • Measurement Timing Function

  • Viscosity value shows continuous change

  • There is an alarm sound prompt when the measurement range is exceeded

  • Standard RTD temperature probe

  • Computer full range linear calibration

  • AC power adapter 100V-240V, good anti-interference

Applications

DV-79A digital viscometer is widely used in starch, glue , latex , solvent-based adhesives , polymer solutions , petroleum , paint coatings , solvents , cosmetics , milk products , medicines , fruit juices, etc.

Pictures

Optional

DV data acquisition and program-controlled analysis software

Special viscosity measurement constant temperature Water Tank

standard solution

micro printer

A container

A, B, C, D rotors

FAQ
QCan this rotational viscometer test the shear force of the foam liquid at the corresponding speed? Speed 0-600 rpm, viscosity range 200-2000mPa · s.
A Hello, Nerun's DV-79A digital Viscometer can meet this speed requirement and viscosity range requirement, but it can only directly test the viscosity value, not the shear force.
QWhich is the water inlet and which is the water outlet of Nirun's DV-79A digital Viscometer test tube?
A Hello, the top is the water outlet and the bottom is the water inlet.
QTo test the shear force of the foam liquid at the corresponding speed, the speed is 0-600 rpm, and the viscosity range is 200-2000mPa · s. Which model is recommended for Nirun.
A Hello, Viscometer can only test viscosity, not shear force directly. To meet the required viscosity range, you can take a look at Nerun's DV-79.
QHow should I choose the right rotor?
A (1) General principle of measurement: high viscosity samples use small volume (G, F) rotor and slow speed, low viscosity samples use large volume (E, F) rotor and fast speed. The percentile scale (torque) of each measurement is between 10% and 90% as the normal value, and the viscosity value measured within this range is the correct value. (2) First estimate the viscosity range of the tested sample, then choose a small volume rotor and a slow speed according to the high viscosity sample, and a large volume rotor and a fast speed for the low viscosity sample. Generally, the rotor is selected first, and then the appropriate speed is selected. For example, when the rotor SP is E, the speed is 750RPM, and the screen directly displays the full scale range of 100mPa.s. When the speed is changed to 75RPM, the full scale range is 1000mPa.s. (3) When the approximate viscosity of the measured sample cannot be estimated, it should be set to a higher viscosity first. Try the rotor from small volume to large volume and the speed from slow to fast. Then each measurement is based on the percentile scale (torque) to judge the rationality of the rotor and speed selection. The percentile scale must be between 10% and 90% of the normal value. If it is not within this range, the Viscometer will issue an alarm to prompt the user to change the speed and rotor. Don't forget to change the rotor. Be sure to change the rotor number SP according to the selected rotor.
QDo you have the instructions for using the Shanghai Nirun DV-79A digital Viscometer?
A Hello, yes, please click the operation manual on the product details page.
QIs there an ISO 12634 standard for digital Viscometers?
A This NIRUN Digital Viscometer will do!
QWhat happens if the Viscometer is outside the measurement range?
A When the current speed and the current rotor measurement range are exceeded, the screen will display and a beeping alarm sound will be issued. When the torque exceeds 100%, the percentage reading and viscosity reading are all EEEE.
QIs the NDJ-8S digital rotational viscometer suitable for measuring industrial wastewater? (Contains various salts, organic matter, garbage permeate, etc.)
A NDJ-8S can not be tested, because the low speed, the sample is prone to sedimentation. It is recommended that DV-79 digital Viscometer with A, B containers.