NIRUN DV-79B Digital Viscometer Medium and High Viscosity
SE
Digital display, Measurement range 1~ 7500000mPa · s, Rotor specifications E, F, G three kinds of rotors, Rotor speed 1~ 800rpm, reproducibility +/- 0.5% F.S, Measurement accuracy +/- 1% F.S.
Digital display, Measurement range 1~ 7500000mPa · s, Rotor specifications E, F, G three kinds of rotors, Rotor speed 1~ 800rpm, reproducibility +/- 0.5% F.S, Measurement accuracy +/- 1% F.S.
Technical advantages
Stepper motor direct drive, no gear, accurate speed
LCD liquid crystal with blue backlight, displaying a large amount of information
It can display the maximum viscosity value that the rotor can measure at the current speed.
Measurement Timing Function
Viscosity value shows continuous change
There is an alarm sound prompt when the measurement range is exceeded
Standard RTD temperature probe
Computer full range linear calibration
AC power adapter 100V-240V, good anti-interference
Applications
DV-79B digital viscometer is widely used in starch, glue , latex , solvent-based adhesives , polymer solutions , petroleum , paint coatings , solvents , cosmetics , milk products , medicines , fruit juices, etc.
Pictures
Optional
DV data acquisition and program-controlled analysis software
Special viscosity measurement constant temperature Water Tank
[Note]Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
QCan this DV79B Viscometer test the viscosity of photosensitive Resin?
A
OK
QHow should I choose the right rotor?
A
(1) General principle of measurement: high viscosity samples use small volume (G, F) rotor and slow speed, low viscosity samples use large volume (E, F) rotor and fast speed. The percentile scale (torque) of each measurement is between 10% and 90% as the normal value, and the viscosity value measured within this range is the correct value.
(2) First estimate the viscosity range of the tested sample, then choose a small volume rotor and a slow speed according to the high viscosity sample, and a large volume rotor and a fast speed for the low viscosity sample. Generally, the rotor is selected first, and then the appropriate speed is selected. For example, when the rotor SP is E, the speed is 750RPM, and the screen directly displays the full scale range of 100mPa.s. When the speed is changed to 75RPM, the full scale range is 1000mPa.s.
(3) When the approximate viscosity of the measured sample cannot be estimated, it should be set to a higher viscosity first. Try the rotor from small volume to large volume and the speed from slow to fast. Then each measurement is based on the percentile scale (torque) to judge the rationality of the rotor and speed selection. The percentile scale must be between 10% and 90% of the normal value. If it is not within this range, the Viscometer will issue an alarm to prompt the user to change the speed and rotor. Don't forget to change the rotor. Be sure to change the rotor number SP according to the selected rotor.
QDoes the NIRUN Digital Viscometer meet the JJG1002 standard?
A
The DV-79B Digital Viscometer can be referred to
Q Can DV-79B Digital Viscometer be timed?
A
Hello, this rotational viscometer comes standard with timing function.
QWhat happens if the Viscometer is outside the measurement range?
A
When the current speed and the current rotor measurement range are exceeded, the screen will display and a beeping alarm sound will be issued. When the torque exceeds 100%, the percentage reading and viscosity reading are all EEEE.
QIs the NDJ-8S digital rotational viscometer suitable for measuring industrial wastewater? (Contains various salts, organic matter, garbage permeate, etc.)
A
NDJ-8S can not be tested, because the low speed, the sample is prone to sedimentation. It is recommended that DV-79 digital Viscometer with A, B containers.