Back

Ion Science PCT-LB-07 VOC Detector Specifications

Search
ION SCIENCE PhoCheck Tiger PCT-LB-07 VOC Detector

Brand:Ion Science

Model:PCT-LB-07

Name:VOC Detector

DetailsContact
Specifications
Minimum resolution 1 ppb or 0.001 mg/m3
Maximum range 20000Pm or 20000mg/m3
Accuracy +/- 5% reading or +/- 1 value
Linearity +/- 5% to 100ppm
intrinsically safe certification ATEX II 1G Ex ia IIC T4 Ga (-15 ℃ ≤ Ta ≤ + 60 ℃), IECEx and CUL
Battery life Lithium battery: up to 24 hours, Charging time 6.5 hours; alkaline battery: 3 × AA, typical 8.5 hours
PID lamp 10.6eV Krypton PID Lamp (Standard) 9.8eV and 11.7eV Lamp Optional
data storage > 120,000 data stores with date and time (if 1 data is stored per second, it can be stored for 33 hours)
communication USB 1.1
calibrate By calibrating tool kit
alarm LED flash and 95dBA 300mm (12 ") speaker
Optional vibration alarm Presets TWA and STEL *; presets include 770 gases *
Flow velocity 220Ml/min with Low Flow rate alarm
Operating Environment Operating temperature: -20~60 ℃, -4~ 140 ° F (non-intrinsically safe); Humidity: 0~99% RH (non-condensing)
Protection level IP65 waterproof design, CE 1180, EMC tested to EN61326-1:2006, EN50270:2006 and CFR47:2008A class
Dimension 340*90*60
TEUs 420×320×97mm(16.5×12.5×3.8”)
Weight 0.72kg(25oz)
Encasement 5kg(176oz)
Packing list
machine X1, manual X1, certificate X1, warranty card X1
After-sales instructions

Warranty days: 365 days

[Note] For after-sales services outside the Chinese mainland, if the instrument requires repair, the shipping and possible associated taxes shall be borne by the sender. Generation address of the manual: http://en.nbchao.com/p/9072/Specification.html , which is only used by users who purchase commodities from the NBCHAO.