GB/T 14352.24-2022 《Methods for chemical analysis of tungsten ores and molybdenum ores—Part 24: Determination of germanium conten—Inductively coupled plasma mass spectrometry》 Related products

This standard specifies a specific method for the determination of germanium content in tungsten and molybdenum ores by Inductance-coupled Plasma Mass Spectrometry. The applicable range for the determination of germanium content is 0.2 μg/g to 300 μg/g, and the detection limit of the method is 0.063 μg/g. The sample is first melted at high temperature by sodiumhydrate and extracted with water. The germanium is present in the solution in the form of germanate. The sample solution is then treated by ion exchange Resin to remove sodium salt and part of molybdenum, and finally analyzed by Inductance-coupled Inductively Coupled Plasma Mass Spectrometer (ICP-MS). Mass Spectrometer performs elemental separation and quantification based on Mass charge ratio, and combines calibrated Linear dispersion to calculate the content of germanium in the sample.

This summary is not the original standard text and is for reference only. For accurate information, please obtain it through official channels.
Status Active
CCS D46
ICS 73.060
Release Date 2022-12-30 00:00:00
Implementation Date 2023-04-01 00:00:00