1. Before the test, first check whether the connection of the lifting beam is reliable and whether it can rotate freely.
2. Wipe the inside of the 4-cup hand-painted cup with gauze or absorbent cotton with the solvent used in the sample, especially the outlet part, and observe it with light until it is wiped clean.
3. Intrude the instrument into the liquid to be tested, lift the lifting beam with one finger and quickly lift the upper plane of the container cup away from the liquid surface, and at the same time start the stopwatch for timing. When the first drop appears in the middle of the sample wire, stop the stopwatch. The time is the full settling time of the measured liquid, that is, the conditional viscosity of the measured liquid.
4. For the conditional viscosity value of the same liquid, the arithmetic mean value should be calculated after three times, but the difference between each measurement should not be greater than 3% of the mean value.
5. After use, the viscometer should be cleaned and wiped clean for storage.