3nh/3nh gloss meter NHG60 is a gloss meter with independent intellectual property rights of 3nh. It is a high-precision gloss meter with an intelligent touch screen. The measuring angle is 60 degrees and the detection spot size is 9*15mm. NHG60 gloss meter fully complies with Chinese national standard GB/T 9754 and international standard ISO 2813. It has a large full touch screen and has the characteristics of convenient use, accurate measurement, stable performance and high-end shape.






1. NHG60 gloss meter has multiple measurement modes of basic measurement, continuous measurement, statistical measurement, and quality control measurement (only host computer), which can meet various usage conditions.
2. It adopts full touch screen operation, which is simple and convenient to operate.
3. The display screen is large, up to 3.5 inches, high-resolution (320*480) full-view display, and the interface image is clear.
4. High-end hardware configuration, incorporating a number of innovative technologies.
5. NHG60 gloss meter complies with the standards ASTM D523, ASTM D2457, GB/T 1381, GB/T 7706, GB/T 8807.
6. Large-capacity storage space, which can store 1000 basic records, 5000 statistical records and 5000 continuous records.
7. With multi-function settings, it can meet the needs of different users.
8. The gloss data can be input manually, which is convenient for customers to operate.
1. The NHG60 gloss meter is a precision optical instrument, please keep and use it properly, and avoid using and storing it in humid, strong electromagnetic interference, strong light, and dusty environments (temperature 20 degrees Celsius, 1 standard atmospheric pressure, humidity 50~70 %RH).
2. The standard plate is a precision optical element, avoid bumping the working surface with sharp objects, avoid soiling the working surface with dirt, and avoid exposing the standard plate under strong light. Regularly clean the working surface of the standard plate with soft sassafras cloth and smoked alcohol, it is best to wipe it only in one direction, do not wipe it back and forth, and make sure that the specific gravity is free of any fine particles. These particles can damage the surface of the standard when cleaning. (Note: Do not use acetone solution )
3. When calibrating, the dust on the working surface should be disposed of in time, and only a clean and intact standard plate can be accurately calibrated.
4. The NHG60 is powered by a built-in lithium battery. If you do not use the NHG60 gloss meter for a long time, please charge the NHG60 gloss meter every 2 weeks to protect the performance of the lithium battery and prolong the life of the lithium battery.
5. It is recommended to return to the factory for cleaning once a year for cleaning the optical objects inside the NHG60 gloss meter. Do not use cleaning tools to insert into the instrument for cleaning, which will cause irreversible damage to the NHG60 gloss meter.