China

RUNQEE G-10 Dual beam of light UV-Visible Spectrophotometer Light spectrum Bandwidth 0.1-5 nm Continuously adjustable

RUNQEE G-10 dual beam of light UV-Visible Spectrophotometer is a fully automatic scanning high-performance UV-Visible Spectrophotometer, Wavelength range of 190~1100nm, G-10 dual beam of light UV-Visible Spectrophotometer with sample temperature control system, sample temperature can be controlled at -5 ℃~ 200 ℃.
Model:

RUNQEE G-10 Dual beam of light UV-Visible SpectrophotometerSPEC

RUNQEE G-10 Dual beam of light UV-Visible SpectrophotometerDetails

RUNQEE G-10 Dual beam of light UV-Visible SpectrophotometerPacking list

SKU
NB014725
Optical inspection system
G-10
Light spectrum Bandwidth
Dual beam of light, continuously adjustable 0.1-5 nm, can customize multi-level Light spectrum Bandwidth (0.1, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0, 5.0nm)
Wavelength range
190~1100nm
Wavelength Repeatability
≤0.1nm
Photometric range
-4.0~4.0A
stray light
±0.0008 Abs
Noise level
±0.0003 Abs
Baseline flatness
≤ +/- 0.0004 Abs/h (at 500nm, after preheating for 2h)
Dimensions
600*505*240mm
Wavelength MPE
+/- 0.1 nm (656.1 nm D2 ), +/- 0.3nm (Full Wavelength range)
Transmittance Precision
±0.2%T
Transmittance Repeatability
≤0.1%T
Baseline drift
±0.0003 Abs
G-10
Dual beam of light UV-Visible Spectrophotometer
RUNQEE G-10 Dual beam of light UV-Visible Spectrophotometer Light spectrum Bandwidth 0.1-5 nm Continuously adjustable Picture

Introduction

This instrument is a dynamic true dual-beam fully automatic scanning high-performance ultraviolet-visible Spectrophotometer controlled by a PC.

The new optical system and circuit design and the continuous adjustment and setting of the spectral bandwidth from 0.1 to 5.0 nm ensure that the instrument has high resolution, low stray light, long-lasting stability and high signal-to-noise ratio, and has powerful software functions, which can meet the needs of many high-level analysts, such as biological, environmental, scientific research and other detection systems.

Functional features

1. Photometric measurement: Transmittance and absorbance can be measured at a fixed wavelength, and 10 wavelength points can also be set at the same time for measurement.

2. Spectral scanning: It can scan the spectral characteristics of samples in any Wavelength range within 190-1100nm, the minimum sampling interval of wavelength is 0.05nm, and the scanning speed is divided into fast, medium and slow, which can derive the spectral curve, peak and valley detection, curve smoothing, map expansion, superposition and a variety of operations.

 3、Quantitative analysis: standard coefficient method, standard control method, dual-wavelength method, three-wavelength and other analysis methods can be applied, and up to 100 can be achieved20Pointed1-4Order curve regression determination.

4、 Kinetic measurements: wavelength points, sampling intervals can be selected, activity calculations can be performed, and can also be performedDNAProtein measurements.

5. Measurement of three-dimensional atlas: the three-dimensional atlas can be made and analyzed for the scanning of some "dynamically changing" samples.

Features

  • Long-lasting stability: The dual-beam dynamic feedback proportional recording metering system and circuit ensure the stability of the baseline and ensure the stability of the instrument for a long time.

  • Ultra-low stray light: Very excellent optical system and professional electronic system ensure that the stray light of the instrument ≤ 0.015%T to make the test data of high-concentration samples more accurate.

  • Automatic 8 spectral bandwidths (0.1, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0, 5.0nm) are continuously adjustable to meet the needs of various sample tests, and the number of spectral bandwidths can also be customized.

  • Full PC operation: the software is rich in content, user-friendly design, and easy to operate. 

  • The lamp source is simple and convenient to replace: the imported socket flange type deuterium lamp and tungsten lamp are adopted, and there is no need for optical path debugging when changing the lamp, and the maintenance is more convenient.

  • Comes with special software: users can connect to the PC to install and use by themselves.

  • * Sample temperature can be monitored.

  • * With the sample temperature control system, the sample temperature can be controlled at -5°C--200°C.

  • The G-10 series dual-beam UV/VIS Spectrophotometer has an open sample chamber and can be equipped with a microcuvette holder, a reflection sample holder, a solid sample holder, a thermostatic Water Bath, and an autosampler for different applications.

Rangqi G-10 dual-beam UV-Vis Spectrophotometer  display

FAQ
QDoes this dual beam of light UV-Visible Spectrophotometer meet the GB/T26810 standard?
A The dual beam of light UV-Visible Spectrophotometer is suitable for GB/T26810, please consult customer service for detailed application definition
Standard