China

BANTE A151-DL Laboratory conductivity /TDS/salinity/resistivity meter

SE
A151-DL laboratory conductivity /TDS/salinity/resistivity meter, professional precision Conductivity Meter, including 4 measurement functions, 7-inch TFT display, the instrument is suitable for measurement of liquid conductivity, total dissolved solids (TDS), salinity and resistivity value, Measurement accuracy: 0.5% F.S.
$616.00/set
PromotionLoading...
Coupon
Model:
Certificate

Bante A151-DL Laboratory conductivity /TDS/salinity/resistivity meterSPEC

Bante A151-DL Laboratory conductivity /TDS/salinity/resistivity meterDetails

Bante A151-DL Laboratory conductivity /TDS/salinity/resistivity meterPacking list

SKU
NB014816
Measuring range
0.05µS/cm~20mS/cm
resolution
0.001, 0.01, 0.1, 1
Conductivity Detector measurement Accuracy
±0.5% F.S
Conductivity Detector calibrated point
1 to 3 point
Calibrating fluid
10µS/cm, 84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm
TDS range
0~10ppt (Max. 20ppt)
TDS measurement Accuracy
±1% F.S.
TDS conversion factor
0.01~ 1.00 (default 0.5)
Salinity range
0~10ppt
Salinity measurement Accuracy
±1% F.S
Salinity measurement mode
Practical salinity (psu), seawater salinity (ppt),% salinity
Resistivity Measurement range
0 ~ 30MZ
Resistivity measurement Accuracy
±1% F.S
Resistivity display resolution
0.01, 0.1, 1
Temperature Measurement range
0~105℃
Temperature measurement Accuracy
±0.5℃
Temperature display resolution
0.1℃
Temperature calibrating point
1 point
Temperature unit
℃/℉
Temperature compensation range
0~100℃
Temperature compensation mode
Linear, nonlinear or pure water compensation
Standardized temperature
20 ℃, 25 ℃, optional
Conductivity Detector cell constant
Two-ring electrode (K = 0.1, 1, 10) or four-ring electrode
Data stability condition
Standard or high precision
Interval reading
10, 30, 60 seconds, 10, 30 minutes or off
Calibrated reminder
1 to 31 days or closed
data storage
1000 groups
Data Interface
USB
Connector
6-Pin MINI Connector
Display
7-Inch TFT LCD
Power
DC12V power adapter
Dimensions
240*200*80mm
Weight
1.7kg
Optional accessories
TP-10K temperature probe, Conductivity Detector standard liquid, CON-0.1 type Conductivity Detector electrode, CON-1 type Conductivity Detector electrode, CON-10 type Conductivity Detector electrode
Packaging Dimensions
300*240*170mm
Encasement Weight
2kg
Temperature compensation method
Manual/Automatic
A151-DL
Laboratory conductivity /TDS/salinity/resistivity meter
BANTE A151-DL Laboratory conductivity /TDS/salinity/resistivity meter Picture

A151 LABOratory Conductivity/TDS/Salinity/Resistivity Meter

Professional precision conductivity meter, including 4 measurement functions, 7-inch TFT display, the meter is suitable for measuring liquid conductivity, total dissolved solids (TDS), salinity and resistivity values, measurement accuracy: 0.5% FS.

Features

• Professional precision conductivity meter with 7-inch TFT display.

• 1 to 5 point key calibration, automatic identification of conductivity standard solution.

• Selectable cell constants to match connected conductivity electrodes and recall calibration data for each sensor.

• Optional linear temperature compensation coefficient, pure water compensation coefficient, normalized temperature and TDS conversion factor.

Precision Conductivity Meter

Precision Conductivity Meter


• Automatic temperature compensation to ensure accurate measurement over the full range.

• Alarm limit, automatically warns if the measured value exceeds the set range.

• Calibration due reminder to remind users to calibrate the meter regularly.

• Automatic end point locking, retaining stable measurement values for easy viewing and recording.

• Interval reading mode, regularly send measured values to PC or printer.

• Password protection to prevent unauthorized calibration and settings.

• Expanded memory can store up to 1000 sets of data.

• USB communication interface is easy to connect external devices.

• One key reset (Reset) automatically restores the instrument to factory default settings


FAQ
QI want to measure water quality, I need to measure conductivity, TDS and salinity, what is the recommendation?
A The BANTE A151-DL Laboratory Conductivity Meter integrates four measurement functions of conductivity, TDS, salinity and resistivity. The 7-inch large screen shows clear display. Measurement accuracy up to 0.5% is especially suitable for water quality testing, such as tap water, drinking water or seawater analysis. It can comprehensively evaluate water purity and salt content.
QDo you have the resistivity meter involved in ASTMD 5682-95?
A The A151-DL resistivity meter will do
Standard