â Adopt high-definition color LCD touch screen with clear display
â Support Chinese and English
â Temperature units optional: ℃ and °F
â Reference temperature 5.0℃, 10.0℃, 15.0℃, 18.0℃, 20.0℃ and 25.0℃ are optional, the default is 25℃
âSupport functions such as automatic shutdown, power-off protection and factory reset
â Support firmware upgrade function, allowing function expansion and personalized requirements
âSupport IP54 protection level
â Comes standard with composite conductivity electrode, independent electrode holder, dust cover and bagged calibration solution
Intelligent operation, automatic identification, batch detection
â Intelligent operating system with functions such as method management, electrode management, data management and user management
â Support instrument self-test and user hierarchical authority management, support password management
â Intelligent identification of end points, supporting automatic reading, scheduled reading, scheduled interval reading, and manual reading
â Automatically identify GB and international standard conductivity solutions, and support 1-5 point conductivity electrode calibration
âSupports calibration reminder and forced calibration functions, supports conductivity standard solution verification and mandatory verification functions
âSupport measurement alarm limit setting
â Within the range, it has functions such as automatic temperature compensation and automatic range switching.
âSupports multiple conductivity compensation modes such as non-compensation, linear, non-linear, pure water compensation, etc.
â Supports multiple salinity compensation modes such as seawater salinity and default salinity.
â Supports automatic frequency switching, and is equipped with a 1.0 constant electrode that can cover the full range of measurements (it is recommended to use a 0.1 constant or 0.01 constant conductivity electrode for measurement of samples below 20 μS/cm)
â Supports connection to an automatic sampler to achieve automatic measurement and batch detection of samples
Data management, information traceability,
â Comply with GLP specifications, support 1000 sets of data storage, review, deletion, statistics, analysis, transmission and printing
â Support GMP management, support GMP operation log review, and achieve complete traceability and tracking of data.
â Equipped with RS-232 interface, supports connection to standard RS-232 serial port printer, and directly prints measurement results. The printing format is optional.
â With USB interface, it can be connected to PC through special communication software to realize data transmission.
â Support U disk reading and writing, support connecting to scanner gun