China

TONGAO NBC-6900D STIP-scan COD Ammonia nitrogen, total phosphorus, total nitrogen turbidity Suspension

Simultaneously detect COD, ammonia nitrogen, total phosphorus, total nitrogen, turbidity, suspended solids several parameters, in line with a variety of national standards, built-in thermal printer, using digestion Colorimetric integrated tube Colorimetric.
Model:
Certificate

TongAo NBC-6900D Desktop STIP-scanSPEC

TongAo NBC-6900D Desktop STIP-scanDetails

TongAo NBC-6900D Desktop STIP-scanPacking list

SKU
NB050418
Measurement parameter
COD, ammonia nitrogen, total phosphorus, total nitrogen, turbidity, suspended solids
measurement error
≤± 5%
Sexual repeability
≤±5%
Optical inspection stability
≤ +/- 0.001A/20min (100,000 hours life)
data storage
30,000 records (including item, date, time, sample ID, etc.)
Linear dispersion quantity
96 Linear dispersion (expandable)
printer
Built-in thermal printer
Display
5-Inch high-definition color LCD screen
batch Volume
16/25 water samples/batch
Colorimetric method
Digestion Colorimetric tube
sample compatibility
16 mm (round), 25mm (round)
communication interface
USB cable, LAN, etc
exception notification
Blank comparison prompt, over-range prompt, etc
operation interface
Membrane push-button operation, Chinese/English (optional)
Power Supply
AC(220V±10%),50Hz
power
5W
protection class
IP30
Environment humidity
≤ 85% non-condensing
Environment temperature
5~40℃
Dimensions
348*250*145mm
instrument Weight
Main unit less than 2.6kg
COD detection method
Method 1: HJ/T399-2007 standard; Method 2: Import detection method
COD Measurement Range
0-20000mg/L;0-20000mg/L
Lower limit of COD detection
3 mg/L;2 mg/L
COD digestion environment
165℃,15min;165℃,20min
COD anti-chlorine interference
1000 mg/L (expandable); 1000 mg/L (expandable)
Ammonia nitrogen detection method
HJ535-2009 "Water quality, determination of ammonia nitrogen, Nessler reagent Spectrophotometric method, spec-trophotometry"
Ammonia nitrogen Measurement Range
0-50 mg/L
Ammonia nitrogen detection lower limit
0.01mg/L
Ammonia nitrogen digestion environment
No need to digest
Total phosphorus detection method
GB11893-89
Total phosphorus Measurement Range
0~20mg/L
Total phosphorus detection limit
0.01mg/L
Total Phosphorus Digestion
125℃,30min
Turbidity detection method
Spectrophotometric method,spec-trophotometry
Turbidity Measurement Range
0~1000NTU
Lower limit of turbidity detection
0.5NTU
Turbidity digestion environment
No need to digest
Suspension detection method
Direct Colorimetry
Suspension Measurement Range
0~500mg/L
Lower limit of suspended solids detection
2mg/L
Suspension digestion environment
No need to digest
TN detection method
Persulfate oxidation
TN Measurement Range
0~250mg/L
TN detection lower limit
0.5 mg/L
TN digestion environment
125℃,30min
NBC-6900D
Desktop STIP-scan
Simultaneously detect COD, ammonia nitrogen, total phosphorus, total nitrogen, turbidity, suspended solids several parameters, in line with a variety of national standards, built-in thermal printer, using digestion Colorimetric integrated tube Colorimetric.
TONGAO NBC-6900D STIP-scan COD Ammonia nitrogen, total phosphorus, total nitrogen turbidity Suspension Picture

Product Introduction

The instrument can quickly determine COD, ammonia nitrogen, total phosphorus, total nitrogen, turbidity, suspended solids, etc. in various water quality, and integrates mature technology in the field of water quality analysis, making the entire analysis system more perfect, simpler and faster operation, and more stable and reliable detection results. It is widely used in emergency monitoring, industrial wastewater, domestic sewage, sewage treatment, chemical industry, food, beverage, textile printing and dyeing, electroplating, papermaking, petroleum metallurgy, river management, scientific research units, colleges and universities and other fields.

Principle of operation:

The digestion colorimetric integrated tube and the multi-channel water quality rapid Tester can be used seamlessly, and are compatible with φ25mm colorimetric tubes.Details of the TR 6900C benchtop water quality Tester Figure 1

● The testing of each parameter is in line with national standards, industry standards or international standards.

● Two built-in COD detection methods

Method 1: HJ/T399-2007 (National Standard)

Method 2: Import testing method

● The standard detection parameters are as follows:

- COD (Prefabricated Reagents)

- COD (Powder Pack Reagent or Custom Reagent)

- Ammonia nitrogen (NH3-N)

- Total Phosphorus (TP)   

- Turbidity - Suspended Solids (SS)

● Strong scalability in the later stage, including but not limited to the following parameters:

- Total nitrogen - phosphate

- Total Chromium - Hexavalent Chromium

- Iron - nickel

- Zinc - Copper

.......

Operating Principle

1. Dissolve colorimetric integrated tube

The digestion colorimetric tube is used for operation, and the digestion of the sample and the reading of the results share a reaction tube, so there is no need to transfer the liquid to the cuvette for testing.

2. Direct display of results

The absorbance of the sample is measured by using a cold light source of a specific wavelength, and the instrument system directly converts the detection result into the sample concentration value through a specific algorithm. 

3. Measure and remove outliers multiple times

Measure the water sample 10 times or more, remove outliers and take an average to reduce the impact of dirt, scratches, fingerprints, or defective glass tubes on the results.

4. Detect abnormal prompts

Indicate detection anomalies, such as blank comparison prompts, result overrun upper limit prompts, etc.


FAQ
QIs the price listed on the website the final cost for overseas purchases?
ANo, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
QWhat is the specific process for purchasing the product?
AFirst, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
QWhat should I do if the instrument I purchase needs repair?
AAfter-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.