Digital display, Measurement range 40~80000000mPa · s, Rotor specifications 2 #, 3 #, 4 #, 5 #, 6 #, 7 #(1 #optional), Rotor speed 0.1~ 200rpm, reproducibility +/- 0.5% F.S, Measurement accuracy +/- 1% F.S, Control system Microcomputer, software system data collection, programming, analysis, temperature sensor PT100 temperature sensor, Encasement aluminum alloy, Rotation speed Control mode continuously variable speed (no gear drive).
Features
Optional DV data acquisition and programmable software measures rheological properties
Displays the shear rate and shear stress
The stepper motor is directly driven, gearless and the speed is accurate
The LCD LCD has a blue backlight and displays a large amount of information
The maximum viscosity value that can be measured by the rotor at the current speed is displayed.
The viscosity value shows a continuous variation
When the measurement range is out, there is an alarm sound
An RTD temperature probe is standard
Bidirectional RS232 or 485 interface as standard
Computerized full-scale linear correction
AC power adapter 100V-240V, good anti-interference
Application
HADV-2+Pro digital viscometer is widely usedInks, glues、latex、Solvent-based adhesives、Polymer solution、petroleum、Paints and coatings、solvent、cosmetics、Milk products、*product、juices and much more.
Product display diagram
Optional
DV data acquisition and program-controlled analysis software
Dedicated viscosity measurement thermostatic sink
Ultra-low viscosity adapters
Small sample adapters
Standard solutions
Micro printers
Externally recirculating glass beaker
Comparison of DV parameters
Model
Measuring range (mPa.s)
Rotor speed
LDV
RDV
HADV
HBDV
DV-1+PRO
1-2 million
100-13 million
200-26 million
800-104000000
0.3-100
DV-2+PRO
1-6 million
100-40 million
200-80 million
800-320000
0.1-200
DV-3+PRO
1-6 million
20-40 million
100-80 million
500-320000
0.1-250
DV-E
1-2 million
100-13 million
200-26 million
800-104000000
0.3-100
Unit: mPa.s, revolutions per minute
Other parameters
Number of rotors
LDV (4 rotors 1, 2, 3, and 4) (rotor 0 is optional)
[Note]Because the manufacturer's packaging may be updated or upgraded, the detailed packaging list shall be subject to the latest standard configuration of the manufacturer.
FAQ
QWhat is the sample volume per test for the HADV-2 + Pro rotational viscometer?
A
Hello, the test sample volume of this Viscometer is 300~ 600ml.
QIs this digital Viscometer compliant with ISO 12634?
A
This digital Viscometer is suitable for ISO12634, please consult customer service for detailed application definition
QShanghai Nirun HADV-2 + Pro Digital Viscometer with which rotor?
A
Hello, HADV-2 + Pro, this Viscometer can be matched with 2/3/4/5/6/7 #of these 6 rotors (1 #can also be matched, but it is not standard and needs to be purchased separately)