All STIP-scan Spectrophotometer
Status
All Active Abolish Replaced Incoming Convert to industry standard
Category
Number Year
Name

YB/T 4217.1-2010 《Coatings of the hot dip galvanizing(zinc-aluminum)sheet.Determination of hexavalent Cr content.Spectrophotometric method》 Related products

YB/T 4217.1-2010 "Hot-dip galvanized (aluminum-zinc) steel platings. Determination of hexavalent chromium content. Spectrophotometric method, spec-trophotometry" specifies the use of diphenylcarbonyl-dihydrazine Spectrophotometric method, spec-trophotometry Determination of hexavalent chromium content in hot-dip galvanized (aluminum-zinc) steel platings principle, scope of application, analytical procedures and results calculation method. This standard applies to the determination range of 0.03 μg/cm ²~ 10.00 μg/cm ² (ie 3 μg/g~ 1000 μg/g) hot-dip galvanized (aluminum-zinc) steel platings in hexavalent chromium content. During the analysis, the sample is leached in boiling water, and hexavalent chromium reacts with diphenylcarbonyl-dihydrazine in an acidic solution to form a purple-red complex. The content of hexavalent chromium is calculated by measuring its Absorbance at a wavelength of 540 nm. This standard refers to "Specifications and Test Methods for Water in Analytical Laboratories" (GB/T 6682) and "Rules for Numerical Revision" (GB/T 8170) to ensure the accuracy and Reliability of analytical results.

This summary is not the original standard text and is for reference only. For accurate information, please obtain it through official channels.
Status Active
CCS (H11) 钢铁与铁合金分析方法
ICS (77.080.01) 黑色金属综合
Release Date 2010-11-10 00:00:00
Implementation Date 2011-03-01 00:00:00
Sort Filter
Extended reading